Philos DR-T Technical Manual
21
PMT Management
Off
; On
6.3.1 Rate Adaptation
Sensor gain
1.0, 1.1, 1.3, 1.4, 1.6, 1.8, 2.0, 2.2, 2.6, 3.0, 3.3, 3.7, 4.0, 4.5,
5.0, 6.0, 7.0, 8.0, 8.5, 10, 11, 12, 14, 16, 18, 20, 22, 24, 28, 32,
35, 40
Sensor threshold
very low; low; mean; high; very high
Rate increase
1, 2, 4, 8 ppm/s
Maximum sensor rate
80...(5)...180 ppm
Rate decrease
0.1, 0.2, 0.4, 0.8 ppm/s
Automatic Sensor Gain
Off
; On
6.3.2 Parameters at Replacement Indication
Basic Rate
Programmed value minus 11%
(in modes DVI(R), DDI(R), DVT(R), DDI/T(R) minus 4.5–11%,
depending on programmed AV Delay)
Magnet Rate
The magnet rate in all modes decreases as shown in the
following table.
Содержание Philos DR-T
Страница 1: ...Philos DR T DDDR Dual Chamber Pulse Generator with Home Monitoring Technical Manual...
Страница 10: ...6 Philos DR T Technical Manual...
Страница 14: ...10 Philos DR T Technical Manual...
Страница 18: ...14 Philos DR T Technical Manual...
Страница 30: ...26 Philos DR T Technical Manual Volume 12 cm3 Dimensions 6 x 44 x 51 mm X Ray Identification VV...
Страница 31: ...Philos DR T Technical Manual 27 7 Order Information Product Name Order Number Philos DR T 331 440...