Adjust the interpupillary distance
(1) Adjust the prism housing along the direction of arrow of
the Fig.4 till the observation is comfortable.
Fig.4
Fig.5
With scale
Screw in
Screw out
Mount and Remove the Optional Eyepiece
Micrometer
(1) Turn and remove the mounting ring from the eyepiece.(Fig.5)
(2) Clean the eyepiece micrometer and mount it to the mounting
with the inscription side downward.
(3) Gently twist the mounting ring with the eyepiece micrometer
into the eyepiece till tightening securely.
(4) To remove the eyepiece micrometer, take down the mounting
ring by twisting and take out of the micrometer, and then wrap it.
Auxiliary
objecive
Fig.6
Fig.7
Fig.8
Use Auxiliary objective
(1) Connect the Auxiliary objective to the thread at the bottom
of the microscope head.
(2) 0.5x Auxiliary objective has long working distance. When use
0.5x auxiliary objective, release the lock screw on the arm by
a 4mm Allen key, move the focus arm up and fixed in the up
screw hole.(Fig 6)
0.75x auxiliary objective no need to remove arm.
Mount PK-mount and photo set
(1) Screw the PK-mount
in the head, M28 screw, tight the
mount.(Fig.7)
(2) Loosen the screw on the PK-mount
, take out the mount of C
type camera
.
(3) Screw the C type mount in the TV photo set.
(4) Install the TV photo set into the PK-mount and lock the screw.
Focusing the TV photo set
(1) Turn the zoom control knob to the maximum magnification,
meanwhile, observe the TV set, adjust until the image is clear.
(2) Rotate the zoom control knob to the min. magnification. If the
image is not clear enough, adjust the part
until the image clear.
(3) Redo the step (1), turn the zoom control knob to the maximum
magnification to check the image is clear or not. If not clear, redo
step(1)and (2), until the image is clear in all the zoom range.