APC/APC Plus Operations Manual
Version 04-11
18
APPENDIX A - Specifications
Illumination Source
....................30 mW laser diode at approximately 670 nm
Particle Size Ranges
..................0.3, 0.5, 1.0, 5.0 microns (APC Plus)
0.5, 5.0 microns (APC)
Flow Rate
....................................0.1 cubic foot per minute ± 10%
Calibration
..................................NIST traceable with polystyrene spheres as per
ASTM F-328 or ASTM F-649
Display
........................................Custom, six-digit LCD display
Data Display
...............................Total particles, particles per cubic foot, or particles per liter.
Temperature Sensor: 10°C to 50°C ± 2°C
Relative Humidity Sensor:
20 to 95% RH ± 10%
Mode
...........................................Actual count, time, date, sample time or hold time
Sampling Mode
..........................Single, automatic, or continuous
Count Time
.................................One second up to eight hours
Hold Time
...................................Two seconds up to 24 hours
M
emory
.......................................Automatic storage of 200 samples
Output
.........................................Serial RS-232C. Baud rate of 9600
Power
..........................................6-volt sealed lead acid battery with charger
Operating Environment
.............Temperature: 10° to 40 °C (50° to 104°F)
Humidity: 0 to 90% (noncondensing)
Storage Environment
.................Temperature: -40°C to 70°C (-40°F to 158°F)
Humidity: 0 to 60%
Dimensions
................................7.7 x 3.8 x 2.6 inches
Weight
.........................................2.5 pounds
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com