WARNINGS
FOR USE WITHIN CELLULAR DATA COVERAGE AREA
Any patient whose life may be put at significant risk by the unavailability of a cellular data system should
not be monitored by the MCOT Patch System.
NOT AN APNEA MONITOR
The MCOT Patch System is not to be used as an apnea monitor.
NOT AN EMERGENCY RESPONSE SERVICE
The MCOT Patch System is not an emergency response service. If you experience any symptoms that
concern you, seek medical help.
DO NOT TAMPER WITH DEVICE
• Do not use the patch if package is tampered or defective.
•
There are no serviceable parts in the Patch/Electrode or Sensor. Disassembling the Patch or Sensor will
void warranty and may alter performance.
• There are no serviceable parts in the Monitor. Disassembling the Monitor will void warranty and may
alter performance.
DO NOT TAMPER WITH MONITOR OR SENSOR BATTERY
The Monitor and Sensor batteries of the MCOT Patch System can present a fire or chemical burn hazard
if mistreated. Do not disassemble, heat, incinerate, or recharge using any device other than the supplied
power cords.
USE POWER CORDS IN SINGULAR OUTLET
A multiple portable socket outlet or extension cord should not be used with the equipment.
USE ONLY MANUFACTURER APPROVED EQUIPMENT
• Do not use any cables, power cords or other accessories other than the ones provided or replaced from
the manufacturer.
•
Usage of cables or accessories other than the ones provided may result in increased Radio Frequency
(RF) emissions or decreased immunity to electromagnetic interference of the system.
SINGLE USE / MULTIPLE USES
The patch is a single use item; do not reuse. All other system components are reusable.
CONNECTION TO COMPONENTS
Only connect the components as described in this manual. Never connect the Patch to any external
electrical item other than the supplied Sensor.
DO NOT STACK
The MCOT Patch System should not be stacked with other equipment. Stacking other equipment on top
of the devices may damage enclosure or inner components.
CAUTIONS
DO NOT TAKE THE MONITOR INTO THE SHOWER
The MCOT Patch System Monitor is water resistant, not waterproof; do not take the Monitor into the shower.
AVOID SPRAYING WATER DIRECTLY ONTO THE SENSOR
The MCOT Patch System Sensor is water resistant, not waterproof. Avoid spraying water directly onto the
Sensor by keeping it away from all direct water flow and deflecting/shielding with body, a hand, or towel,
when showering.
DO NOT LIE DIRECTLY ON THE SENSOR
Do not lie directly on the Sensor as damage may result. Avoid sleeping directly on top of the Sensor.
PATCH STORAGE TEMPERATURE AND SHELF-LIFE
The Patch's storage temperature requirement is narrower than the rest of the components and
temperature limits should be taken into account for storage. Shelf like of 18 months from Date of
Manufacturer has been tested within the range of +5ºC to +27ºC. Storage of Patch outside these limits
may affect longevity of shelf life.
DISPOSAL
The patch is single use. Dispose of the Patch properly in accordance with your local ordinances or
instructions of your prescribing physician. The Patch may need to be recycled in accordance with local laws.
All other items provided are reusable and should be returned at the end of your monitoring period.
Specifications
Sensor
General
Channels - 2 channels
Resolution - 12 bits
Sample Rate - 250 Hz
Environmental
IPX Rating When Inserted into Patch/Electrode - IPX4 Resistant to water splashes from any direction"
Operating Temperature * - 10°C to +45°C
Non-Operating Temperature - -20°C to +70°C
Operating Humidity - 10% to 95% (non-condensing)
Non-Operating Humidity - 5% to 95% (non-condensing)
Physical
Size - 2” x 1.6” x 0.36”
Weight - .069 oz. (19g)
Connection to USB - Female Micro USB
Connection to Patch/Electrode - Custom 8 terminal connector
Battery Type - Rechargeable Lithium Ion 3.7 V 500 mA-Hr