Chapter 5: As-Needed Maintenance
Clean the Tip Priming Trough and Priming Plate
1.
Remove the tip priming trough from the left rear of the instrument’s microplate
carrier (see below).
2.
Wash the tip priming trough and priming plate in hot, soapy water. Use a small
brush or cloth to clean the corners of the trough and plate.
3.
To decontaminate
,
soak the trough and plate in a container of 0.5% sodium
hypochlorite
or
70% isopropyl alcohol for 20 to 30 minutes.
4.
If decontaminating in bleach solution, remove the trough and plate, and
thoroughly rinse with DI water.
If decontaminating with alcohol, remove the trough and plate and let them air-
dry.
5.
Discard the used gloves and cloths using a biohazard trash bag and an
approved biohazard container.
Figure 33: Tip priming trough and priming plate
tip priming
trough
BioTek Instruments, Inc.
Содержание Synergy HTX
Страница 1: ...Operator s Manual Multi Mode Microplate Reader Synergy HTX ...
Страница 2: ......
Страница 12: ...viii Preface Revision History Rev Date Changes A 7 2014 Initial release BioTek Instruments Inc ...
Страница 22: ...xviii Preface BioTek Instruments Inc ...
Страница 28: ...6 Chapter 1 Introduction BioTek Instruments Inc ...
Страница 31: ...1 Unpack and Inspect the Reader 9 Figure 1 Unpacking the reader accessories box Synergy HTX Operator s Manual ...
Страница 52: ...30 Chapter 2 Installation BioTek Instruments Inc ...
Страница 72: ...50 Chapter 3 Getting Started BioTek Instruments Inc ...
Страница 102: ...80 Chapter 4 Preventive Maintenance BioTek Instruments Inc ...
Страница 166: ......
Страница 174: ...150 Appendix A Specifications BioTek Instruments Inc ...