Injection System
The tubing and injectors should be cleaned at least every three months. See
Clean the
Dispense Tubes and Injectors
on page
for instructions.
Inspect the injection system daily for leaks, preferably immediately after priming and
whenever plumbing changes have been made.
If a syringe is leaking, it may need to be replaced. See
on page
for instructions.
Dispense Module
The dispense module sits on top of the reader and pumps fluid from the reagent bottles
to injectors located inside the instrument. Fluid is injected into one well at a time. The
injectors support plate types from 6- to 384-well plates.
Figure 3-2: Dispense module components
1
Two 250-µL syringes draw fluid from the supply bottles.
2
Inlet tubes transport fluid from the supply bottles to the syringes. These tubes
are short pieces of opaque PTFE (Teflon) tubing connected to stainless-steel
probes on one end and threaded fittings on the other end.
3
Valves switch the syringe flow from the inlet tubes to the outlet tubes.
Synergy H1 Operator's Manual
Internal Components | 33
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...