224 | Chapter 5: Qualification
50 µL Solution
Mix 45 mL of DI or dH2O water with 5 mL of the
5 µL Solution
.
100 µL Solution
Mix 40 mL of DI or dH2O water with 2 mL of the
5 µL Solution
. (Use this solution
to test the RAD 8-to-1 cassette.)
1536 Solution
Mix 5 mL of 70% isopropyl alcohol with 3 mL of the
5 µL Solution
and 35 mL of DI
H2O.
Perform the Peri-Pump Precision and Accuracy Tests
Prerequisite:
l
Gather the
l
Prepare the
.
l
Make a copy of the applicable worksheets. Find them on the operator's manual USB stick
in the Qualification chapter PDF.
Procedure:
1. Install the cassette to be tested.
2. Turn on the MultiFlo FX and make sure the cassette type setting is correct.
3. Turn on the balance.
4. Fill a beaker or other vessel with the test solution.
5. Define a
Protocol
and save it for reuse. Set the parameters based on the desired
test volume:
n
A predefined protocol for the
1536 Test
is shipped with the instrument and
installed on your PC when you install the LHC:
QC_P-1536_DISP_TEST
(P2
protocols are designed for the Secondary Peri-pump).
l
Add a dispense step to the protocol:
.
l
Set the
Dispense Volume
to match the Test:
Test
Volume
1 µL
1 µL
MultiFlo™ FX Multi-Mode Dispenser
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......