108 | Chapter 3: Operation
1536-well plate
Volume (µL) 3-3000
The 32-tube manifold flow rates do
not have minimum volumes.
The µL/sec/well for each type of
manifold, Small Bore (SB) and Large
Bore (LB), is shown:
Rate
SB
LB
1
56
125
2
58
150
3
60
162
4
62
174
5
64
187
The default rate is 3.
Low-density-plate manifolds
Like the standard manifolds described above, for these plate-type-specific
manifolds the flow rate is defined as
µL/sec/well
and the minimum volumes are
cumulative, i.e. any rate can be used to dispense Rate 5's minimum.
Manifold
Flow/Minimum
Volume
Rate 1
Rate 2
Rate 3
Rate 4
Rate 5
6-Well
Flow
560
836
1116
1400
1680
Min Vol
40
80
200
240
320
12-Well
Flow
420
627
837
1050
1260
Min Vol
30
60
140
160
220
24-Well
Flow
280
418
558
700
840
Min Vol
20
40
100
120
160
48-Well
Flow
279
372
560
600
800
Min Vol
40
80
200
240
320
Wash Parameters Table
See Protocol Parameters Tables on page 103
for Peri-pump and Syringe
Dispenser parameters.
See Plate Types Table on page 9
for default Z-axis values
or dispense heights.
Minimally, a wash step includes an aspirate step followed by a dispense step.
Select and define each option to customize the parameters for your assay.
MultiFlo™ FX Multi-Mode Dispenser
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......