6
| Chapter 1: Introduction
Optical Specifications for single-wavelength kinetic measurements with read
intervals of less than 12 seconds:
Accuracy:
± 2.0% ± 0.010 OD from 0.000 to 2.500 OD @ 405 nm
Linearity:
± 2.0% from 0.000 to 2.500 OD @ 405 nm
Repeatability:
± 1.0% ± 0.010 OD from 0.000 to 2.500 OD @ 405 nm
The
plate can be read at
6-second
intervals (rapid mode) with
wavelengths higher than 400 nm.
Incubator/Ultraviolet Model: ELx808IU
The following specifications apply only to standard 96-well,
flat- or round-bottom microplates.
•
Wavelength Range
: 340 to 900 nm
•
Filters
: 10 nm half-bandwidth interference filters. User-accessible filter wheel.
Up to 6 filters may be installed on the instrument at one time. Filters supplied:
340, 405, 450, 490, 630 nm.
•
Absorbance Measurement Range
: 0.000 to 4.000 OD for 400 to 900 nm range,
0.000 to 3.000 OD for 340 to 400 nm range
Optical specifications for the 400 to 900 nm range (12-second read in normal/
regular read mode):
Accuracy, Linearity, Repeatability
: Same as Standard Model.
Optical specifications for the 340 to 400 nm range (12-second read in normal/
regular mode):
Accuracy:
± 1.0% ± 0.010 OD from 0.000 to 2.000 OD @ 340 nm
Linearity:
± 1.0% from 0.000 to 2.000 OD @ 340 nm
Repeatability:
± 1.0% ± 0.005 OD from 0.000 to 2.000 OD @ 340 nm
•
Incubation
: The following specifications apply to a 96-well sealed plate with
200
μ
L of liquid in all wells:
¾
Temperature Control
: Temperature controlled to 50
°
C
¾
Temperature Variation
: ± 0.50
°
C @ 37
°
C with the plate sealed
BioTek Instruments, Inc.
Содержание ELx808
Страница 1: ...Operator s Manual Absorbance Microplate Reader ELx808 ...
Страница 2: ......
Страница 30: ...8 Chapter 1 Introduction BioTek Instruments Inc ...
Страница 100: ...78 Chapter 3 Operation BioTek Instruments Inc ...
Страница 136: ...114 Chapter 5 Preventive Maintenance BioTek Instruments Inc ...
Страница 144: ...122 Chapter 6 Error Codes BioTek Instruments Inc ...
Страница 146: ...124 Appendix A Sample Reports Samples with Calls on Matrix Report BioTek Instruments Inc ...
Страница 147: ...Curve Fit Report 125 Curve Fit Report ELx808 Operator s Manual ...
Страница 148: ...126 Appendix A Sample Reports BioTek Instruments Inc Samples with Calls on Column Report ...
Страница 149: ...Column Report without Samples 127 Column Report without Samples ELx808 Operator s Manual ...
Страница 150: ...128 Appendix A Sample Reports Panel Report BioTek Instruments Inc ...
Страница 151: ...Assay Detail Report 129 Assay Detail Report ELx808 Operator s Manual ...
Страница 172: ...150 Appendix C Adjusting the Power Input Voltage Setting BioTek Instruments Inc ...