Cleaning:

The various components of the chamber can be cleaned as follows:

1.  The closure assembly and power/sensor connector do not come into contact with the specimen and should only be cleaned 

with a damp cloth or alcohol wipe.  DO NOT immerse either one in liquid.
2.  Coverslips and gaskets should be hand cleaned and short cycle autoclaved before use.  Do not use alcohol on gaskets.
3.  The white plastic (Delrin) portion of the FCS2 will be in fluid communication with the specimen and can be autoclaved 

at 120°C for 15 minutes.

High Numeric Aperture Use:

If you are using high numeric aperture (immersion) objectives with high N.A. dishes, it will be necessary to regulate the 
temperature of the objective as well.  This is due to the fact that the optical coupling medium, oil, glycerin or water also has a 
thermal coupling effect.  In this case, a Bioptechs, Inc. Objective Heater and an Objective Heater Controller will be necessary 
for uniform temperature across the field.

Objective Heater Description:

The Bioptechs Objective Heater is an adjustable, thin film, size and shape compliant, heating band and temperature sensor 
assembly attached to a linear translator.  This device attaches to the objective and plugs into a separate Temperature Controller, 
which regulates the objective temperature within 0.1°C of SETPOINT.

Continual Perfusion with the Micro-Perfusion Pump:

Bioptechs recommends the FCS2 Micro-Perfusion Pump due to its smooth flow profile in slow flow applications.  If dual flow 
is needed there is also a PC interface board and program that enables alternate control of two pumps to regulate the acceleration 
and run time of each.  The Micro-Perfusion Pump Controller (PC interface) is ideal for changing concentrations of media.

Содержание FCS2

Страница 1: ... of the information displayed When the controller is turned on the default condition is to display the setpoint value indicated by the red light adjacent to the word Setpoint The button selects one of a sequence of display options This button can be pressed at any time and has no effect on the temperature regulation functions of the controller Display Lights indicate the source of the value displa...

Страница 2: ...ogramming or cycling operating temperatures Controller Operation 1 Turn the unit on using the main on off switch located on the rear panel The Chamber need not be plugged in at this time All further adjustments are made on the front panel 2 Set the display switch to the SET position Insert the enclosed screwdriver into the hole marked Setpoint and adjust the controller to the desired setpoint 3 Se...

Страница 3: ...e and press around the perimeter creating a seal Perfuse some media through the perfusion port producing a bead on the surface of the slide to displace air trapped in the line Lower the coverslip with cells onto the bead until it is resting on the gasket Place the closure assembly on top of this stack while aligning the black electrical connector with the oval slot in the chamber top Maintain a ge...

Страница 4: ...ll This is due to the fact that the optical coupling medium oil glycerin or water also has a thermal coupling effect In this case a Bioptechs Inc Objective Heater and an Objective Heater Controller will be necessary for uniform temperature across the field Objective Heater Description The Bioptechs Objective Heater is an adjustable thin film size and shape compliant heating band and temperature se...

Отзывы: