Operation
32
P/N 128097 Rev. 00
TurboVap 96 User’s Manual
Biotage AB
Elapsed Time Display
The Value display blinks “- -” when the position has been stopped or
has not been started.
If evaporation has started, the Value display shows the number of
minutes since the start of the evaporation. The maximum elapsed
time is determined by the Gas Time setting, which can range from 1
to 99 minutes. If the Gas Time is set to 00, the elapsed time
displays “
” after 99 minutes.
The
Set
button and the up and down buttons have no effect on the
elapsed time display.
If you press the
Select
button, the Cycle Readout light is lit and the
display rotates as described below.
Cycle Readout
When Cycle Readout is selected, the Value and Units displays
show the current conditions for each parameter in the following
sequence:
•
Position 1 - Gas Time,
•
Position 1 - Plate Temp
•
Position 1 - Flow Rate
•
Position 1 - Elapsed Time
•
Position 2 - Gas Time
•
Position 2 - Plate Temp
•
Position 2 - Flow Rate
•
Position 2 - Elapsed Time
After displaying the Position 2 – Elapsed Time, the display cycles
back to Position 1 - Gas Time.
If evaporation is in process, the
current
plate temperature is
displayed, not the
set
Plate Temperature.
If the current temperature is +/- 10
C from the set Plate
Temperature, the displayed temperature blinks.
If only one position is running, the display only shows the settings
for that position. If both positions are running or stopped, the
settings for both positions are displayed.
If you press the
Select
button when Cycle Readout is selected, the
display shows the previously set Gas Time for Position 2, and the
Position 2 – Gas Time LED will light. (The up, down, and Set
buttons have no effect when Cycle Readout is selected.)