13
Evaporate
Manual Mode
Figure 20 displays the Manual mode view. The display shows
that the left side is idle (the orange
Start
button is active) and
the right side is running an evaporation (the blue
Pause
button
and the red
Stop
button are active).
To perform a Manual mode run, enter desired evaporation
parameters (gas flow, gas temperature, plate temperature,
and plate height). To initiate the run, press the
Start
button.
To terminate the run, press the
Stop
button. To pause the run,
press the
Pause
button.
1
Gas flow
The current and set gas flow
2
Gas temperature
The current and set gas temperature
3
Plate temperature The current and set plate temperature
4
Plate height
The current and set plate height. Press to
move the plate up or down.
5
Start
Press to start the evaporation
6
Pause
Press to pause the evaporation
7
Stop
Press to stop the evaporation
Figure 20.
The Manual Mode View. In this figure the manifold settings are
synchronized.
Time Mode
Figure 21 displays the Time mode view. The display shows that
the left side is ready to start (the orange
Start
button is active)
and the right side is running an evaporation (the blue
Pause
button and the red
Stop
button are active).
To perform a Time mode run, enter desired evaporation
parameters, including time. To start the run, press the
Start
button. The progress bar displays the remaining and elapsed
time for the current evaporation. To change a setting, see
"Change a Setting in Manual or Time Mode" on page 14.
To terminate the run before it reaches the set end time, press
Stop
. To pause the run, press
Pause
. The pause will temporarily
turn off the gas flow.
1
Gas flow
The current and set gas flow
2
Gas temperature
The current and set gas temperature
3
Plate temperature The current and set plate temperature
4
Plate height
The current and set plate height. Press to
move the plate up or down.
5
Time
The set evaporation time in minutes
6
Start
Press to start the evaporation
7
Elapsed/
Remaining time
The progress bar shows remaining and
elapsed time of the current evaporation.
8
Pause
Press to pause the evaporation
9
Stop
Press to stop the evaporation
Figure 21.
The Time Mode View. In this figure the manifold settings are
synchronized.