System Overview
E-mail Functionality
N
With the system connected to an e-mail server, it is possible for
the users to:
»
Receive e-mail notifications
when user interaction
is required for their runs, e.g. when a rack has to
be replaced, a solvent needs to be replenished,
a waste reservoir needs to be emptied, or a leak is
detected by the optional Biotage Solvent Detector.
»
Automatically receive the report by e-mail
when
their purification run has been completed.
»
Send reports and logs from the system by e-mail.
Figure 10.
E-mail notifications and auto send of reports are enabled and
disabled individually in each user account.
N
Audible Alarm
When the audible alarm is enabled in the system settings,
a warning will sound if an error has occurred.
Biotage
®
Safety Valve
The optional safety valve (P/N 417115SP) enables you to use
columns or combinations (e.g. column and dry load vessel)
with a combined column volume larger than 0.8 L and up to
a maximum of 3.1 L.
The safety valve is designed to open at 8 bar and discharge
fluid until the pressure drops below 8 bar.
Spektra Software License (Optional)
The following features are included in the Spektra software
license (P/N SEL-SPK) that is sold separately:
»
λ
-All detection.
The system uses average
absorbance within a user-defined wavelength
range for collection and fractionation.
»
UV Baseline Correction.
When the UV baseline
correction option is turned on, the gradient run is
preceded by a light absorbance detection phase.
During this phase, the light absorbance of the used
solvents (A and B) is measured for the whole detector
range using either the initial gradient mix to the end
mix A/B or the initial gradient mix to 100% solvent B.
The measurement results in a baseline that is subtracted
from the detector signal during the gradient run.
»
Absorbance spectrum.
The absorbance spectrum for
the whole detector range is displayed during the run.
»
Tool for spectrum analysis.
The analysis view shows the
chromatogram and the absorbance spectrum for the whole
detector range for a selected point in the chromatogram.
RFID Upgrade (Optional)
The RFID Upgrade Kit (P/N 417504SP) includes an RFID reader as
part of the system, and enables automatic recognition of Selekt
collection racks making setup easier and adding the possibility
to find results corresponding to fractions collected in a specific
rack (see “RFID Tags
Functions requiring this upgrade are marked with a superscript
R in this user manual.
Network Upgrade (Optional)
With the Network Upgrade Kit (P/N 417501SP), it is possible to
enable network share and/or e-mail functionality (see “Network
Share
” on page 3 and “E-mail Functionality
” above).
Functions requiring this upgrade are marked with a superscript
N in this user manual.
N
Requires the Network Upgrade Kit (P/N 417501SP).
R
Requires the RFID Upgrade Kit (P/N 417504SP).