411829-L, Maintenance
April 2012
Page 5-12
5
7. Pull out the UV lamp and disconnect it from the power cable; see Figure 5-14.
Figure 5-14. Disconnect the UV Lamp
NOTE
Do no touch the UV lamp glass.
8. Mount a new UV lamp (P/N 09830 for all internal Isolera detectors except UV-VIS and P/N
412970 for UV-VIS). Ensure the power cable is positioned so that it does not get in contact with
the fans at the left side of the UV lamp.
9. If using a UV-VIS detector, replace the tungsten lamp (P/N 412970):
a. Loosen the screw holding the tungsten lamp in position using the wrench delivered with
the system.
b. Pull out the tungsten lamp and disconnect it from the power cable; see Figure 5-15.
c. Mount a new tungsten lamp. Ensure the lamp is fully inserted and the power cable is
positioned so that it does not get in contact with the fans at the left side of the UV lamp.
Figure 5-15. Disconnect the Tungsten Lamp
10. When you are done, put the service panel back in place.
11. Reconnect the system and turn it on.
12. When the main menu appears, log into the System mode (see page 3-1).
13. Ensure that the detector flow cell is clean (see page 5-3) and then perform an intensity
calibration:
a. Select the
Maintenance
tab in the right-hand panel.
b. Press
Calibrate
in the
UV Detector
field. Read and follow the instructions that appear on
the screen.
Press here and
disconnect the
UV lamp
Locking screw
Press here
and disconnect
the lamp