411829-L, Operation
April 2012
Page 4-13
4
4.4.3 Create a Method Using the Method Wizard
The method wizard will guide you step-by-step through the setup of a purification with a gradient
including up to three steps using two solvents.
1. Select the
Method
tab in the right-hand panel.
2. Press
Wizard...
. The method wizard opens.
Figure 4-8. Method Wizard
3. Read and follow the instructions that appear on the screen. (See Table 4-1 on page 4-7 for
guidance in selecting rack type.)
4. If you want to save the method and reuse it in the future:
a. Press
Save to Editor
.
b. Enter a unique method name by selecting the
Method Name
text box at the
Parameters
tab.
c. Press
Save
. If your user account is password-protected, the
Password
dialog opens.
Enter your password, using the keypad, and press
OK
.
5. To run a purification using the method, press
Run...
.
6. Prepare the system as described on page 4-15.
7. Proceed with step 5 in “Run a Purification” on page 4-19.
4.4.4 Gradient Optimization
Linear gradients can be automatically optimized by converting them to step gradients which
reduces solvent use.
NOTE
Gradients that are optimized on an Isolera Prime, Isolera One, or Isolera
Four system cannot be run on an Isolera LS system due to the different
system volumes used for the calculation, and vice versa.
Clear separation between the peaks is required to receive a relevant result.
1. Select the
Results
tab in the right-hand panel.
2. Select the run at the
Result Selection
tab. For information on how to find the desired run, see
“Search for Records” on page 4-28.
3. Press
Optimize...
. The
Gradient Optimization
dialog opens.