 Flash 400 Quick Guide | Page 1

Biotage

®

 Flash 400 Quick Guide

Control Panel

Item

Description

PG-01

Pump pressure gauge. Monitors the solvent pump outlet 

pressure.

PG-02

Cartridge compression pressure gauge. Monitors the 

pressure inside the radial compression module.

MV-01

Solvent select valve. Selects which of the two solvent 

inlets (A or B) that will be used.

MV-06

Fraction select valve. Selects which of the two outlets 

(1 or 2) the elution stream will exit through.

MV-07

Cartridge compression valve.

Opens the line between the radial compression 

chamber and the radial compression fluid reservoir, 

which compresses the cartridge if the fluid reservoir is 

pressurized. PG-02 and PG-03 readings will equalize.
Closes the valve.

MV-08

Solvent pump valve. Turns the solvent pump on or off 

(O or I) by opening or closing the line between the air 

supply and the pump.

MV-09

Reservoir pressure valve.

Pressurizes the radial compression fluid reservoir by 

connecting it to the nitrogen supply. 
Used when checking that the pressure in the radial 

compression fluid reservoir can be maintained.
Vents the radial compression fluid reservoir by 

opening it to the atmosphere. The exhaust port is 

located on the side connection panel.

MV-11

Pump flow rate control. Regulates the compressed 

air pressure to the solvent pump. Turn the knob 

counterclockwise to increase the solvent flow rate and 

clockwise to decrease it.

Safety Positions

Before introducing compressed air or nitrogen into the system, 
ensure that the following valves are in their safety position.

Valve

Safety Position

MV-11

Clockwise until stopped

MV-08

O (closed)

MV-09

 (vent)

MV-07

 (open/compress)

Radial Compression Fluid Reservoir

Item

Function

PG-03

Reservoir pressure gauge. Monitors the radial 

compression fluid reservoir pressure.

Side Connection Panel

Item

Description

Solvent inlets A and B.

Fraction outlets 1 and 2.

MV-10

Radial compression drain valve. 

Closes the valve.

Drains the fluid from the radial compression 

chamber. Used before removing the cartridge.

MV-05

Vacuum break valve, bottom. 

 

Closes the valve.
Breaks any vacuum that develops when the cartridge 

is in use. Used when preparing to remove a cartridge 

or when testing if the cartridge seal adapter’s O-ring 

has sealed properly. 

Compressed air inlet for the solvent pump.

Nitrogen inlet. Used to pressurize the radial compression 

fluid reservoir and the radial compression module.

Exhaust port for when venting the radial compression 

fluid reservoir using the MV-09 valve.
Safety relief port for the radial compression module. The 

module is rated at 170 psig (11.7 bar) maximum pressure. 

Top Head Assembly

Item

Description

MV-02

Three-way injection valve. Selects the solvent or sample 

inlet. During cartridge removal, compressed gas may be 

blown through this valve and through the cartridge to 

remove solvent.

MV-03

Cartridge compression vent valve. Open the valve to 

purge air from the radial compression chamber when 

compressing the cartridge.

MV-04

Vacuum break valve, top. Open the valve to break any 

vacuum that develops when the cartridge is in use. Used 

when preparing to remove a cartridge or testing if the 

cartridge seal adapter’s O-ring has sealed properly.

Figure 1. The cartridge compression vent valve (MV-03) and the top 

vacuum break valve (MV-04) closed (left image) and opened (right image).