Site Requirements and Specifications
2
Performance Specifications
Specification Conditions
ASTM: “Evaporative Light Scattering Detectors Used in Liquid
Chromatography”.
ASTM drift tests require a temperature change below 2 °C/h
(3.6 °F/h) over one hour period. Our published drift specification is
based on these conditions. Larger ambient temperature changes will
result in larger drift. Better drift performance depends on better
control of the temperature fluctuations. To realize the highest
performance, minimize the frequency and the amplitude of the
temperature changes to below 1 °C/h (1.8 °F/h). Turbulences around
one minute or less can be ignored. ASTM measurements require that
the detector should be turned on enough time before start of testing.
Table 3
Reference Conditions
Gas Flow
1.6 SLM
Neb Temperature
40 °C
Evaporator Temperature
40 °C
PMT Gain
1
Data Rate
40 Hz
Smoothing
30
Light Source Intensity
100 %