background image

14

monitoring for at least fifteen minutes.  Until
the minimum monitoring time has elapsed,
the STEL screen will show an "X" where
the reading should be.

  

  

After fifteen minutes have elapsed, the
STEL screen will show the appropriate
numerical value.

  

   

Press the MODE button again to display
TWA (Time Weighted Average) exposure
levels.  Time Weighted Average or TWA
values are calculated by taking the sum of
exposure to a particular toxic gas in the
current operating session in terms of parts-
per-million-hours and dividing by an eight-
hour period.  It is not possible to compute a
toxic gas TWA until the Toxi Plus has been
monitoring for at least one hour.  Until the
minimum monitoring time has elapsed, the
TWA screen will show an "X" where the
reading should be.

  

  

After an hour has elapsed, the TWA screen
will show the appropriate numerical value.

  

  

Pressing the MODE button again to return
current gas readings.
Regardless of operating mode selection,
anytime the Toxi Plus is on, it is
remembering the peak readings of all
gases measured, and is calculating both
long term Time Weighted Averages and
Short Term Exposure Levels (for toxic gas
reading detectors) and will go into alarm
whenever appropriate.

2.2.4 Changing operating modes

It’s possible to change operating modes
while the instrument is in use without
affecting the on-going calculations used to
determine Peak, STEL and TWA alarms.
Simply slide the belt clip at the back of the
instrument out of the way to expose the

four mini-push-buttons.  Then
simultaneously press and hold the "+" and
"-" buttons until the word “MODE” is
displayed.  Each time the operating mode
is changed the display screen will briefly
indicate which operating mode has been
selected.

Note: Shifting modes or otherwise
reprogramming the instrument should
only be done by employees who are
authorized to do so.

Figure 2.2.  Toxi Plus with belt clip

removed exposing the four keypad

push-buttons

2.3

Toxi Plus batteries

The Toxi Plus uses a set of three AAA
disposable alkaline batteries.  A fresh set of
batteries should power the instrument for
up to 1,500 hours of operation.

2.3.1 Low battery voltage
alarms

When it is time to replace the batteries a
"

B

" will appear in the lower left hand corner

of the display screen, and the audible
alarm will “beep” once per minute until the
batteries are replaced.  Once the “B” is
shown, the user should have approximately
one to two weeks of useful battery life
remaining.

2.3.2 Battery replacement

Toxi Plus batteries should be replaced as
soon as possible after a low battery ("

B

”)

message has been noticed.
Use the following procedure to change the
batteries.

Содержание Toxi Plus

Страница 1: ...0 Reference Manual Toxi Plus Single Sensor Gas Detector 651 South Main St Middletown CT 06457 860 344 1079 FAX 860 344 1068 04AUG2004 P N 13 044 Version 3 3...

Страница 2: ...1...

Страница 3: ...ONS IT IS ESSENTIAL THAT THE INSTRUCTIONS IN THIS MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED AVERTISSEMENT LIRE ATTENTIVEMENT LES INSTRUCTIONS AVANT DE METTRE EN MARCHE Toxi Plus Operation Manual Co...

Страница 4: ...mote sample draw assembly 10 1 9 4 Weather cover 10 1 9 5 Training video 10 1 10 Toxi Plus design components 10 1 11 Toxi Plus accessories 11 Chapter 2 Basic operation 12 2 1 Overview of Toxi Plus ope...

Страница 5: ...Re setting the calibration gas concentration 25 4 3 Setting the security beep interval 26 Chapter 5 Trouble Shooting and Repair 27 5 1 Replacing Toxi Plus sensor 27 5 2 Troubleshooting 27 5 2 1 Unit w...

Страница 6: ...to day operation including fully automatic calibration adjustment through use of the single on off Mode button The Toxi Plus is Classified by Underwriters Laboratories Inc and the Canadian Standards...

Страница 7: ...oxi Plus instruments should be checked periodically with known concentration calibration gas Failure to check accuracy can lead to inaccurate and potentially dangerous readings 5 A sensor that cannot...

Страница 8: ...7...

Страница 9: ...us is designed to detect is determined by the sensor The type of sensor must be specified at the time the instrument is purchased The types of sensors currently available include substance specific el...

Страница 10: ...With the alarms latched once an alarm occurs both visual and audible alarms continue to sound even after the atmospheric hazard has cleared The instrument must be manually reset to silence the alarms...

Страница 11: ...nd H2S The sensor selected must be specified at the time of purchase 1 9 2 Vibrator alarm An internally mounted vibrating alarm is optionally available for use in high noise or other environments requ...

Страница 12: ...trols all normal day to day operations including Auto Calibration adjustment Accurate electrochemical sensor Easy to read LCD display with backlight Loud two tone audible alarm Optional built in vibra...

Страница 13: ...ator alarm will briefly activate During start up the LCD display will also show several messages or screens in sequence The first screen displays the software version number Note The software version...

Страница 14: ...ormal operation of the instrument resumes 2 2 3 Technician mode Technician Mode provides access to all advanced functions and displays When the instrument is turned on in the Technician Mode following...

Страница 15: ...nging operating modes It s possible to change operating modes while the instrument is in use without affecting the on going calculations used to determine Peak STEL and TWA alarms Simply slide the bel...

Страница 16: ...moisture barrier filter Normal air movements are enough to carry the sample to the sensor The sensor reacts to changes in the concentration of the gas being measured Values are quickly updated and dis...

Страница 17: ...kits and long lasting cylinders of test gas specifically developed for easy Toxi Plus calibration Customers are urged to use Biosystems calibration materials when calibrating their Toxi Plus detector...

Страница 18: ...ach is usually to use a direct reading substance specific sensor If hydrogen sulfide is known to be potentially present the sensor selected should be specifically for the detection of H2S and calibrat...

Страница 19: ...esh air If the Toxi Plus is operated in either the Basic or Technician mode readings should equal 0 PPM If the instrument is operated in the Text Only mode all readings should indicate conditions are...

Страница 20: ...initiating auto calibration procedures 3 Make sure the instrument is located in an area where the air is known to be fresh 4 Press the MODE button 3 times within two seconds This will wake up the inst...

Страница 21: ...as it is screwed into the cylinder of gas 3 The instrument will display the sensor readings as they rise When they have stabilized at their highest value the instrument will note the response and if...

Страница 22: ...ly which verifies that the zero values have not been changed The instrument will then return to normal operation 3 5 2 Span calibration using keypad buttons Span calibration procedures using buttons o...

Страница 23: ...it may indicate that the sensor is unresponsive or dead and requires replacement 11 When span calibration has been completed press and hold the Cal button down until the LCD screen indicates that cal...

Страница 24: ...o sound even after the atmospheric hazard has cleared Press the MODE button to reset the instrument silence the alarms and restore normal operation Procedures for latching Toxi Plus alarms are given i...

Страница 25: ...ode as discussed in Section 2 2 4 then turn the instrument off 2 Turn the instrument on while pressing and holding down the Alarm button This will place the instrument in the Alarm Adjust Mode 3 Adjus...

Страница 26: ...ry When gas is applied to the sensor during auto calibration the instrument automatically adjusts the sensitivity of the carbon monoxide to produce readings that match this 50 PPM concentration Sensor...

Страница 27: ...ocedure to turn the security beep on 1 Make sure the Toxi Plus detector is in the Technician mode as discussed in Section 2 2 4 then turn the instrument off 2 Turn the instrument on while pressing and...

Страница 28: ...g and repair procedures that can be done in the field Caution The following repair procedures should only be performed by authorized personnel 5 2 1 Unit will not turn on Possible causes Battery disch...

Страница 29: ...ny calibration gas fittings take the instrument to fresh air and allow readings to stabilize Do a manual fresh air zero adjustment using buttons on the instrument keypad as discussed in Section 3 5 1...

Страница 30: ...ood example of an acutely toxic substance which is immediately lethal at relatively low concentrations Exposure to a 1 000 PPM parts per million concentration of H2S in air produces rapid paralysis of...

Страница 31: ...osure limits which are higher than the eight hour TWA The STEL is the maximum average concentration to which an unprotected worker may be exposed in any fifteen minute interval during the day During t...

Страница 32: ...rms should be set If OSHA has not determined a ceiling value for greatest safety the Toxi Plus ceiling alarm should be set at the same value as the STEL alarm If OSHA has not determined a STEL value t...

Страница 33: ...alibration easy in this case it is necessary to use special calibration Zero Contaminant gas This gas cylinder Biosystems part number 9039 is used in conjunction with the sample draw calibration adapt...

Страница 34: ...H2S Sulfur dioxide SO2 CO 100 10 100 25 0 H2S 10 100 350 100 0 SO2 10 20 65 15 100 NO 30 0 25 6 0 NO2 15 20 60 15 120 Cl2 10 20 40 10 5 H2 60 5 60 15 0 HCN 15 0 40 10 50 HCl 3 0 5 1 0 NH3 0 0 0 Ethy...

Страница 35: ...e instrument is in service Toxi Oxy Pro 2 years from date of purchase Toxi Oxy Limited 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Mighty Tox 90 days...

Отзывы: