1. Error Messages
The InBody S10 displays the following error messages to warn the user of the problems run
into during operations and to guide the users to take steps. The following are the most
common error messages and the steps to handle the corresponding errors.
A. “Data entered is outside of accepted range. Please RE-ENTER.”
This message appears when the value for age or height
of the examinee is out of the permissible range for these
data. Check your entry again. As for the permissible
recommended input range of each data, refer to the
“Chapter 2, section 8 : Personal Profile.”
B. “Please check electrode connection.”
This message fires up when the posture of the examinee
is not appropriate or the examinee’s palms or soles are too
dry or have too much hard skin, making it impossible to start
the test. Correct the posture of examinee or wipe examinee’s
hands and feet with an electrolyte tissue before re-measuring
impedance.
Содержание inbody s10
Страница 18: ...7 When you use the touch type electrodes When you use the adhesive type electrodes ...
Страница 25: ...14 Thermal printer cables are connected in the same manner as shown in the picture ...
Страница 52: ...40 4 ICW ICW ECW History 5 TBW FFM History 6 Weight Muscle PBF History ...
Страница 62: ...50 Chapter3 Setup Establishment 1 Setup 2 Setup Menu 3 DATABASE ...
Страница 70: ...56 Chapter4 Problems Solutions 1 Error Messages 2 Troubleshooting 3 Frequently Asked Question FAQs ...
Страница 77: ...Chapter5 Consumables 1 Consumables 2 Options ...
Страница 84: ...69 IB Battery 110 connection diagram When use InBody S10 bag When use InBody S10 cart ...
Страница 85: ...70 Appendix 1 More about the InBody S10 2 Specifications 3 Customer Service Information ...