8.3 

UV-lamp replacement

. Replacement must be performed only by qualified and spe-

cially trained  personnel.  UV lamp  replacement is  necessary  after  lamp  stops func-
tioning or at the end of manufacturer specified lifetime. Use the indicator in the centre 
of recirculator (fig. 4/2 or 5/8) to check UV lamps inside the recirculator. If the indicator 
is alight while the switch is ON, then the UV lamps are functioning. If it is not, replace 
the lamp, single lamp in case of 

UVR-M

, both - in case of 

UVR-Mi

 

Caution! 

Do  not  switch  on  the  unit  with  the  cover  removed  or  without  filters!  UV 
lamp must be covered at all times during operation. Otherwise, UV lamp 
can expose the operator and other people to dangerous levels of UV light. 

8.4 

Fuse replacement

. Disconnect the unit from the mains and unscrew the fuse holder 

cap (fig. 4/3 or 5/10). Check the fuse and replace, if necessary, with a correct one: 

Model, version 

Voltage  

Fuse 

UVR-M

: V.4AA, V.4AB, V.4A3 

220 V  

M 1 A 

UVR-M

: V.4AC 

120 V  

M 1.6 A 

UVR-Mi

: V.3AA, V.3AB, V.3A3 

220 V  

M 2 A 

 

Fuse type M 

– time delay Medium. 

8.5 

Decommissioning of the unit. 

The user is responsible for the decontamination of 

the unit before decommissioning. Dispose of the appliance as electronic equipment 
in accordance with the relevant national laws. 

9. 

Storage and transportation 

9.1 

Store and transport the unit at ambient temperatures between -

20°C and +60°C and 

maximum relative humidity of 80%. 

9.2 

After transportation or storage and before connecting it to the electric circuit, keep the 
unit under room temperature for 2-3 hrs.  

9.3 

For extended storage, the unit does not require special procedures. 

10.  Warranty 

10.1  The  Manufacturer  guarantees  the  compliance  of  the  unit  with  the  requirements  of 

Specifications, provided the Customer follows the operation, storage and transporta-
tion instructions. 

10.2  The warranted service life of the unit from the date of its delivery to the Customer is 

24 months. For extended warranty, see 

10.5

10.3  Warranty covers only the units transported in the original package. 
10.4  If  any  manufacturing  defects  are  discovered  by  the  Customer,  an  unsatisfactory 

equipment report shall be compiled, certified and sent to the local distributor address. 
To obtain the claim form, visit 

Technical support

 page on our website at link below. 

 

 

Содержание UVR-M

Страница 1: ......

Страница 2: ...services we would like to hear from you Please send all feedback to Manufacturer SIA Biosan Ratsupites iela 7 k 2 Riga LV 1067 Latvia Phone 371 674 261 37 Fax 371 674 281 01 https biosan lv Marketing...

Страница 3: ...special attention to sections marked by this symbol Caution Do not switch on the unit with the cover removed or without filters UV lamp must be covered at all times during operation Otherwise UV lamp...

Страница 4: ...specific at mospheres Do not operate the unit if it is faulty or has been installed incorrectly Do not use outside laboratory rooms 2 4 Biological safety The user is responsible to carry out appropri...

Страница 5: ...t damage Warranty covers only the units transported in the original package 4 2 Complete set Package contents 4 2 1 Standard set UVR M UVR Mi UV Cleaner Recirculator 1 pce Spare dust filters 2 pcs Spa...

Страница 6: ...ta given by the manufacturer so lamp replacement or air recircu lation time extension by 12 is recommended in order to achieve the de sired air cleaning result Figure 4 UVR M overview 5 1 Working with...

Страница 7: ...the next parameter 5 2 3 3If no key is pressed for more than 10 s the changes will be saved 5 2 4 The unit can operate in automatic or manual modes 5 2 4 1Manual mode Press the Run Stop key fig 5 7 to...

Страница 8: ...nd maximum rel ative humidity 80 for temperatures up to 31 C decreasing linearly to 50 relative humidity at 40 C Biosan is committed to a continuous programme of improvement and reserves the right to...

Страница 9: ...rating integrity check If the unit follow the procedure described in section 5 Op eration then no additional checks are required 8 2 Cleaning and disinfection 8 2 1 Use mild soap and water with a soft...

Страница 10: ...y Medium 8 5 Decommissioning of the unit The user is responsible for the decontamination of the unit before decommissioning Dispose of the appliance as electronic equipment in accordance with the rele...

Страница 11: ...and retain for your records Model Serial number Date of sale UVR M UVR Mi UV Cleaner Recirculator 10 8 Production date Production date is placed in the serial number on the label of the unit Serial n...

Страница 12: ...SIA Biosan R tsup tes iela 7 k 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 https biosan lv Edition 3 4 01 October of 2021...

Отзывы: