7
4. Getting started
4.1.
Unpacking.
Remove packing materials carefully and retain them for future shipment
or storage of the unit. Examine the unit carefully for any damage incurred during
transit. The warranty does not cover in-transit damage. Warranty covers only the units
transported in the original package.
4.2.
Complete set.
Package contents:
4.2.1. Standard set:
-
DEN-1 / DEN-1B
Densitometer, suspension turbidity detector ........................ 1 pce.
-
A-16
adapter for Ø16 mm tubes ....................................................................... 1 pce.
-
AA batteries (for
DEN-1B
only) ........................................................................ 3 pcs.
-
External power supply ..................................................................................... 1 pce.
-
User instructions, declaration of conformity .................................................... 1 copy
4.2.2. Optional accessories, on request:
-
A-12
adapter for Ø12 mm tubes ....................................................................... 1 pce.
-
CKG16
calibration kit for Ø16 mm glass tubes .................................................. 1 set
-
Calibration kit for Ø18 mm glass tubes............................................................... 1 set
-
Calibration kit for Ø12 mm glass tubes (for
DEN-1B
only) ................................. 1 set
-
Glass sample tubes without lid, ........................................................ 1 set of 78 pcs.
Figure 1. View from below
4.3.
Setup
.
-
(for
DEN-1B
) Battery installation. Pull back the pin on the battery compartment (fig.
1/2) on the underside and open lid. Insert batteries as shown on the scheme inside
and close the lid.
-
Place the unit on horizontal even working surface.
-
Connect the external power supply unit into the socket at the rear side of the unit
(fig. 3/2) and position the unit for an easy access to the external power supply and
the power switch.
Note.
For
DEN-1B
, connecting external power supply is optional when the bat-
teries are in use.
-
Remove the protective film from the display.
4.4.
Factory calibration
. The device is pre-calibrated at the factory for operation with the
glass tubes 16 mm in external diameter (see the label on the bottom side of the unit,
fig. 1/1) at temperature range from +15°C to +25°C and saves calibration data when
being switched off.
Note.
Recalibrate the unit before using the tubes that are different from factory
calibrated (e.g. with different outer diameter, bottom shape or different
material such as plastic). See the
Calibration
section of this manual.
Содержание BS-050102-AAF