3.7 Clearing All Values
If you are sure that you want to permanently remove all stored values, hold down the button (the
instrument must have been switched off before hand) until the “CL” appears and then release the
button. If you do not want to clear the values, press the button. To permanently clear the memory,
press the button while “CL” is flashing. Individual values cannot be cleared.
3.8 Discontinuing a Measurement
If it is necessary to interrupt a blood pressure measurement for any reason (e.g. the patient feels unwell), the
button can be pressed at any time. The device then immediately lowers the cuff pressure automatically.
4. Software Functions
This unit can be used in connection with your personal computer (PC) running the PC Link Blood Pressure Analyzer
Software. The software will allow a capacity of monitoring 80 patients, each with 1000 data (Note: overuse will lower
system efficiency). The memory data can be transferred to the PC by connecting the monitor via USB cable with your
PC.
NOTE:
The software does not work with Mac Computers.
NOTE:
The USB cable can be purchased at any electronics store. You will need USB cable Mini-B 5 pin. USB versions
1.0 or 2.0 will work on this device.
System Requirements:
This software is compatible with Microsoft
®
Windows
®
XP / Vista / Windows 7 and supports 32 and 64 bit operating
systems.
To download the software go to www.
bios
diagnostics.com. Go to the support page and download “3MS1-4K Link
Software 3.2.5.E”. Follow the directions to download the software.
Once installation is complete connect the monitor via USB cable with the PC. Three horizontal bars will appear on the
display and last for 3 seconds.
The bars will then flash to indicate that the connection between computer and device is successfully made. As long
as the cable is plugged in, the bars will keep flashing and the buttons are disabled.
During the connection, the device is completely controlled by the computer. Please refer to the ‘help’ file in the
software for detailed instructions.
5. Error Messages / Malfunctions
If an error occurs during a measurement, a long beep followed by two short beeps is generated and the LCD display
the corresponding error code.
Error
Possible Cause
Remedy
Err 1
No pulse has been detected.
Ensure that the cuff is being worn correctly, and
that you have your arm at the heart level.
Err 2
Unnatural pressure impulses influence the mea-
surement result. Reason: The arm was moved
during the measurement.
Avoid unnecessary movement or talking.
Err 3
The inflation of the cuff takes too long. The cuff
is not correctly seated.
Ensure that the cuff is being worn correctly.