14
12
、
The common trouble and trouble shooting
The LCD display window
lights on and displays the relative error code when
the error/trouble appears.
Form 3
If the centrifuge still can not work after removing the above several troubles,
please contact with our company or service stations designated by our company.
13
、
Transportation and storage
13.1 Transportation
·You should use the wooden and paper case during long-distance transportation. Put
centrifuge with dust cap into the case, and fill with foam or plastic shock absorption
materials. It is forbidden to be hit, inverted, rolled and soaked by rain and snow.
·You can directly move the centrifuge in room, but avoid big vibration, hitting and
inversion.
13.2 Storage
· If it is leaving unused for long time, you should open the door lid and store it in
ventilated, dry and clean room where there is no corrosive, flammable and explosive
matters.
· As the technology continuously renews, please consult us if there is something
unconformable with this instruction manual.
Error
code
Reasons
Solution
E4
The door lid
isn’t closed
The door lid isn’t closed during run.
Check and Close door lid
The limit switch of door lid lock is not
in place or damaged.
Check and replace limit
switch
E6
Overspeed
Motor drive module fault
Check and replace
E7
Motor runs
without speed.
No power for motor.
Check and replace motor
drive module
Motor shaft is blocked.
Check and replace motor
Содержание TDZ4-WS