PRODUCT SHEET
www.biopac.com
BIOPAC Hardware |
BioNomadix Series
| Page 9 - 26
Updated: 8.18.2020
Table 3: BioNomadix Combo Pairs
– See Table 1 for Biopotentials, Table 2 for Dual Transducer and Table 4 for Accelerometer
BioNomadix
BN-RSPEC
BN-PPGED
BN-NICO
BN-DYNEMG
Signal type:
RSP plus ECG
PPG plus EDA
Z and dZ/dt
Dynamometry plus EMG
Excitation:
--
--
Type: Alternating current
sine wave
Current: 1 mA rms
Frequency: 50 kHz
--
BandlimitsMax:
Factory preset:
Filter Options:
Respiration (CH A):
see BN-RSP2 spec
ECG (CH B) :
see BN-ECG2 spec
Both: DC to 10 Hz:
PPG: 0.5 Hz to 3 Hz
EDA: DC to 3 Hz
Both: DC, 0.5 Hz HP, 3 or 10 Hz LP
EDA: 1 Hz LP
Both: DC to 50 Hz
*
Both: DC to 50 Hz
*
DC, 1, 3, 5, 50 Hz
*
LP
*
Units shipped before
11/2016 are bandlimited
to 10 Hz.
Dyn: DC 100 Hz
Dyn: DC to 10 Hz
Dyn: DC, 3 Hz, 10 Hz, or
100 Hz LP
EMG: see BN-EMG2
specs
Notch filter:
50/60 Hz user-controlled switch; typically not required
—factory preset OFF
.
See Appendix for additional hardware-specific
output options.
Resolution:
see BN-RSP2 and BN-ECG2
specs
PPG: FSR/4096; (4.88 mV)
EDA: 0.012 µS (min step)
Z: nominally ~0.05
Ω
(rms) at 10 Hz BW
dZ/dt: ~0.0075
Ω/sec
(rms) at 10 Hz BW
Dyn: 35 micro kg-f/cm2
(0.0005 psi) (rms)
EMG: see BN-EMG specs
Signal range:
see BN-RSP2 and BN-ECG2
specs
PPG: ±10 V (at output)
EDA: 0 to 50 µS;
excitation
: 0.5 V
constant V
Z: 5 to 100 Ω (mag)
dZ/dt:
±10 Ω/sec
Dyn: 0
– 1.055 kg-f/cm2
EMG: up to 10 mV P-P
Output Voltage range:
± 10 V (receiver output)
Operating time:
72-90 hours
24 hours
24 hours
75 hours
Included strap:
137 cm - BN-STRAP137
33 cm - BN-STRAP33
137 cm - BN-STRAP137
33 cm - BN-STRAP-33
Input:
CH A: BN-RESP-XDCR
CH B: BN-ELxx-LEAD3
CH A: BN-PULSE-XDCR or BN-
PULSEEAR-XDR
CH B: BN-EDA-LEAD2 or BN-EDA25-
LEAD2
2 x BN-EL50-LEAD4
(or 2 x BN-EL50-LEAD2)
CH A: BN-CLENCH-XDCR
CH B: BN-ELxx-LEAD3