Indication
Error description
E04: Temperature Error Unknown Barcode
The environmental temperature is above or below the operating
range of the analyzer.
solution
Move to an area in the acceptable temperature range for the
analyzer and perform the test. Do not heat or cool the analyzer
artificially.
E05: Communication Error
Communication between the analyzer and barcode reader or
printer has failed.
solution
Reconnect the analyzer and the external device. If the error
continues after turning the analyzer ON/OFF, please contact BioNote,
Inc.
E06: Total hemoglobin out of range
The measured total hemoglobin is outside the range of 7 to 23g/dL.
solution
This error occurs with specimens known to have total hemoglobin in
the abnormal range. If the error continues after turning the analyzer
ON/OFF, please contact BioNote, Inc.
Result: Invalid
The test is invalid.
solution
Retest with a new test device and a new specimen. If the error
continues after turning the analyzer ON/OFF, please contact BioNote,
Inc.