Error: Operation module Piston Encoder Error
While operating the cartridge in the M10 Module there was
a problem with the operation of the vertical piston motor.
Solution
If error occurs, the tests will be canceled and the door of
M10 Module will be opened. Proceed the test with a new
cartridge. If the symptom recurs, contact the supplier.
Error: Check the connected module information
Error occurring in M10 communication control module.
Solution
If the Connect number is displayed differently from the Total
number, turn off the power of M10 Console and M10 Module
and check if the cable and gender are properly connected.
Error: Operation module operation error
Error occurring in M10 motor operation module.
Solution
Turn off the power of M10 Console and M10 Module and
check if the cable and gender are properly connected.
Error: Optic module operation error
Error in M10 Optical Measurement Module.
Solution
Turn off the power of M10 Console and M10 Module and
check if the cable and gender are properly connected.
Error: Console Socket error
Error occurring in M10 socket communication console.
Solution
Turn Off the power of M10 Console and M10 Module and
check if the cable and gender are properly connected.