3. Non-Averaging set and canceled:
You may select the unwanted or doubtful test results, to exclude them from the average calculation, after your
every test. After you finished your test in the test mode, you can exclude the unwanted result. Keep pressing the
Right butten until you see the symbol “NO AVG” in the top right cornetr of the screen. By keeping the Right
button pressed and pressing the main button simultaneously to confirm the value as “Non average”.
Now this value is flagged by the symbol “NO AVG” in the up per right corner of the screen.
If you want to cancel this non-averaging setting, you have to stay in the test mode. First keep pressing the Right
button until the “NO AVG” symbol changes to “AVG” in the up per right corner of the screen. To confirm this
original status you again simultaneously press the Main button to confirm the “AVG” status. Now the value will be
stored as a regular value and included in the average values.
Recalling Test Result and Average
NOTE
- The average function is related to the time setting. You must set the time right and have enough time
intervals on the basis of current inquiry time, to make sure that the average test results will show. Use the
14-day average for example; if your current inquiry time is 2010/1/30, then that you must check that you
have tested blood glucose in the past 14 days before 1/30, including today. If not, the 14-day average will
show no figure.
- The Non-averaging/re-averaging function only works in the test mode and when you get the value. In the
MEM mode, you can not perform this action.
37
5. Reading right after test: If you just finish the test, press the Main button to enter the MEM (memory) screen and
review the latest test result.
Recalling Test Result and Average
4. Quick Searching: If you want to see all values automatically displayed in sequence, first enter the memory mode.
Then keep pressing the Right or Left button for 2 seconds. (Right button is for searching the tests the latest value
to the oldest one; by using the Left button you can scroll from the oldest to the latest one). Anytime you want to
stop just release the pressed button. It will stop and display the current displayed value. You can use this way to
find a particular reading you want to check.
NOTE
Your Control Solution results will be automatically non-averaged without selecting, but still can be recalled.
The Control Solution Test results will show with “CS” symbol under memory screen.
PRECAUTION
-
You have to set the time and date to activate the average function.
- The " " , " " results, the Control Solution results and the test result made out of normal temperature
range <50°F
(10°C),
>104°F (40°C)
are not calculated in the average.
3. Non-Averaging set and canceled:
You may select the unwanted or doubtful test results, to exclude them from the average calculation, after your
every test. After you finished your test in the test mode, you can exclude the unwanted result. Keep pressing the
Right butten until you see the symbol “NO AVG” in the top right cornetr of the screen. By keeping the Right
button pressed and pressing the main button simultaneously to confirm the value as “Non average”.
Now this value is flagged by the symbol “NO AVG” in the up per right corner of the screen.
If you want to cancel this non-averaging setting, you have to stay in the test mode. First keep pressing the Right
button until the “NO AVG” symbol changes to “AVG” in the up per right corner of the screen. To confirm this
original status you again simultaneously press the Main button to confirm the “AVG” status. Now the value will be
stored as a regular value and included in the average values.