Getting Ready for Testing

Performing a Blood Glucose Test

Before performing a blood glucose test, prepare the items below :

Rightest

 Blood Glucose Meter 

 Blood Glucose Test Strips 

 

   (Please check the expiration date on the test strip vial. Do not use 

expired test strips)

Rightest

 GD500 Lancing device

- Sterile lancet
- Alcohol swab (optional)

GM232B

GM232B

Rightest

Plunger

Base

Lancet carrier

Release button

Depth adjustable cap

1. Hold the adjustable depth cap in one hand while holding the base in the other 

hand. Lift the cap until a gap appears between the cap and base.

2. Pull the two parts in opposite directions to remove the adjustable depth cap.

3. Insert a new disposable lancet firmly into lancet carrier.

4. Twist off and set aside the protective cover of the disposable lancet.

5. Replace the depth adjustable cap.

6. Choose a depth of penetration by rotating the top portion of the depth adjustable 

cap until your desired setting is visible in the window. Settings are based on skin 
type "          " for soft or thin skin; "          " for average skin; "          " for thick or 
calloused skin. 

1

2

3

6

4

5

Rightest

All manuals and user guides at all-guides.com

all-guides.com

Содержание GM232B

Страница 1: ...40242 Taiwan R O C Made in Taiwan USA Service Center BIONIME USA CORPORATION 1450 E Spruce Street Bldg B Ontario CA 91761 USA Tel 888 481 8485 www gediabetes com Rightest blood glucose monitoring syst...

Страница 2: ...All manuals and user guides at all guides com...

Страница 3: ...sults to help you better manage your diabetes The Rightest Blood Glucose Monitoring System GM232B is designed for in vitro diagnostic use only for self testing by a single user outside the body Testin...

Страница 4: ...or diagnose diabetes mellitus If the Rightest Blood Glucose Meter GM232B and Test Strips are exposed to a substantial change in temperature please wait 45 minutes before measurement Follow all environ...

Страница 5: ...M232B and Lancing Device are for single patient use Do not share them with anyone including other family members All parts of the kit are considered biohazardous and can potentially transmit infectiou...

Страница 6: ...ding Test Results and Messages Rightest Rightest Rightest Rightest Quality Control About Quality Control Testing Performing a Quality Control Test Understanding Control Test Results Test Memory Recall...

Страница 7: ...Package Insert 5 Blood Glucose Meter GM232B with 2 AAA batteries installed 6 Blood Glucose Test strips GM232B 10 25 pcs 7 8 Lancing Device 9 Clear Cap 10 Disposable Sterile Lancets 10 pcs 11 Carrying...

Страница 8: ...Window Test Strip Port Right Button Product name Battery Cover Meter Serial Number Meter Battery The Rightest Blood Glucose Meter GM232B The Rightest Blood Glucose Meter GM232B Rightest GM232B All man...

Страница 9: ...n an error occurs Current date under time mode or testing date under memory mode Warns when the operational temperature limit is exceeded during testing Indicates the time in 12H format Displays curre...

Страница 10: ...When you open a new vial of Blood Glucose Test Strips record the date on the vial Discard the vial of test strips after 3 months from opening Store the Blood Glucose Test Strip between 39 86 F or 4 3...

Страница 11: ...scard a used battery Risk of explosion if battery is replaced by an incorrect type 4 The Blood Glucose Meter 5 Press any button to exit the self test and enter Setting Mode Rightest all symbols will a...

Страница 12: ...ain button to turn on the Rightest Blood Glucose Meter Hold down the main button for 7 seconds During this time the screen will go blank until you hear a beep After the beep the meter will turn on int...

Страница 13: ...confirm Once confirmed the Hour Setting will appear Once the hour is confirmed the Minute Setting will appear Once the minute is confirmed the Volume Setting will appear 5 6 4a 4b NOTE If you do not...

Страница 14: ...the Bluetooth In Time mode press and hold the Right or Left button for two seconds The blinking symbol will switch to the symbol with a beep If Volume is turned on 3 Manual turn on the Bluetooth In T...

Страница 15: ...and firmly stops Removing the Rightest Blood Glucose Test Strip GM232B 1 Hold the Rightest Blood Glucose Test Strip GM232B as shown 2 Rotate the Blood Glucose Test Strip counterclockwise and pull up...

Страница 16: ...e holding the base in the other hand Lift the cap until a gap appears between the cap and base 2 Pull the two parts in opposite directions to remove the adjustable depth cap 3 Insert a new disposable...

Страница 17: ...The will automatically detect the Code number on the strip You do not have to check the Code number on the meter display and strip vial Rightest Blood Glucose Meter GM232B 12 A blood drop icon will ap...

Страница 18: ...oo soon If this occurs please repeat the test with a new test strip Record the opening date of a new test strip vial Discard the vial of test strips 3 months after opening Always keep the metal contac...

Страница 19: ...ion NOTE If no button pressed the marker will be set blank automatically after 15 seconds Pressing for a few seconds Releasing button Alternative site testing palm or forearm blood sampling To perform...

Страница 20: ...ystem that is known to be well maintained and monitored by a healthcare provider use the clear cap provided with the Rightest GD500 Lancing Device when testing Make sure your blood sample covers the w...

Страница 21: ...symptoms that are not consistent with your blood glucose test results and you have made sure to follow all instructions in this manual contact your healthcare professional immediately GM232B GM232B R...

Страница 22: ...er control solutions please contact the GE Diabetes Customer Support Center Rightest Control Solution GC550 Level 4 Level 2 toll free at 1 866 613 7085 When should a Quality Control Test Be Performed...

Страница 23: ...st Strip from the vial and close the vial cap immediately 2 Insert the Blood Glucose Test Strip horizontally not vertically with the colored view window facing up into the test strip port 3 While the...

Страница 24: ...error message and Do not drip the control solution on to the sample port of the test strip directly The reagent on the strip might leak into the bottle of control solution and may cause the degenerati...

Страница 25: ...the test was performed using the The control solution test result will not be used when recalling test average results Rightest Control Solution GC550 1 Press the Main button to switch the screen to...

Страница 26: ...Mode you will not be able to perform the exclusion The Lo Hi results control solution test results and test results under abnormal temperature 50 F 10 C 104 F 40 C will be excluded from average calcu...

Страница 27: ...and lancing device should be cleaned before being disinfected CaviWipes Disinfecting Towelettes has been tested and may be used to clean and disinfect the meter and lancing device Users may purchase t...

Страница 28: ...the typical life of the meter 5 years 2 Allow to remain wet for 2 minutes 1 CAUTION Users should wash their hands thoroughly with soap and water after handling the meter lancing device or test strips...

Страница 29: ...er will not function and the symbol will blink 3 If the Rightest Blood Glucose Meter and Test Strips are exposed to a substantial change in temperature please wait 45 minutes before measurement GM232B...

Страница 30: ...atteries 2 Wait for 5 minutes and reinsert the batteries as described as instructed on page 16 Meter Activation and Changing the Battery The meter should work normally after finishing the above steps...

Страница 31: ...ately low results High concentrations of Ascorbic acid Vitamin C 5 mg dL Cholesterol 600mg dL and Uric acid 10 mg dL may result in inaccurate test results Not for neonatal use For over the counter use...

Страница 32: ...been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC rules These limits are designed to provide reasonable protection against harmful interferenc...

Страница 33: ...s 50 104 F 10 40 C 10 90 500 blood glucose test results with date and time Turns off automatically 2 minutes after last user action To turn off manually press themain button for 3 seconds This device...

Страница 34: ...warranty card GE is a trademark of General Electric Company and is used under license by Bionime Corporation No 100 Sec 2 Daqing St South Dist Taichung City 40242 Taiwan This warranty does not apply...

Страница 35: ...tomer service toll free at 888 481 8485 We will make every effort to provide assistance and solutions for you Description of used symbols References 1 American Diabetes Association Standard of Medical...

Страница 36: ...6833 Klaus Vlbg Austria Disposable Sterile Lancets Manufacturer SteriLance Medical SuZhou Inc No 68 LiTangHe RD XiangCheng SuZhou JiangSu 215133 P R China Product complies with Medical Device Directiv...

Страница 37: ...ose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucos...

Страница 38: ...ose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucos...

Страница 39: ...ose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucos...

Страница 40: ...ose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucos...

Страница 41: ...Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucose Insulin Me...

Страница 42: ...ose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucos...

Страница 43: ...ose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucos...

Страница 44: ...S S Blood Glucose nsulin Medication I Blood Glucose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Warranty Card Distributor name...

Страница 45: ...r GM232B 3 The sole obligation of Bionime under this warranty shall be to replace any defective Blood Glucose Meter GM232B No other warranties express or implied are made Bionime shall not be responsi...

Отзывы: