background image

Surgical Technique

Addendum to the Vanguard

®

 

Complete Knee System

Vanguard

®

 

CR Stem Housing 

Knees  •

 

Hips  •  Extremities  •  Cement  and  Accessories  •  PMI

®

  •  Technology

Содержание Vanguard CR

Страница 1: ...Surgical Technique Addendum to the Vanguard Complete Knee System Vanguard CR Stem Housing Knees Hips Extremities Cement and Accessories PMI Technology ...

Страница 2: ...uction 1 Setting Medial Lateral Position 1 Reaming 2 Distal Cut 2 Drilling for Femoral Lugs Optional 3 Trial Reduction 3 Implant Assembly 4 Ordering Information Implants 5 Instrumentation 6 Table of Contents ...

Страница 3: ...g that the correct side is facing out Place the stem housing reamer guide flush with the anterior cortex and distal femur Determine the appropriate M L position Figure 1a Pin the stem housing guide in place using 1 8in pins through the anterior plate Remove the M L wing and snap the stem housing reamer guide bushing into place Figure 2 This surgical technique is utilized by Edward McPherson M D Bi...

Страница 4: ...ntil diaphyseal cortical contact is achieved Figure 3a For diameters under 18mm the stem housing reamer must be used to complete bone preparation Figure 3 Distal Cut Re cut the distal femur 5mm to accommodate the stem housing attachment plate using the slot on the anterior side of the Vanguard stem housing reamer guide Figure 4 Figure 3 Figure 4 Figure 3a ...

Страница 5: ... posterior bone Tighten the thumb screw of the canal guide Figure 5 Use a in drill through the holes in the distal peg guide If posterior augments are required a trial can be threaded onto the distal peg guide Trial Reduction Select the correct size trial CR stem housing and adjust the position of the bolts by aligning the black line to the correct size as indicated by the color coded dot Figure 6...

Страница 6: ...ocking screw and tighten Bolt the stem housing to the femoral component using the two supplied bolts Figures 7 and 8 Optional If you desire femoral pegs select the 5mm Vanguard augment pegs and attach the CR stem housing construct to the femoral component Figure 7 Figure 8 ...

Страница 7: ... Vanguard CR Stem Housing Vanguard CR Stem Housing Vanguard CR Stem Housing Vanguard CR Stem Housing Vanguard CR Stem Housing Vanguard CR Stem Housing Vanguard CR Stem Housing Vanguard CR Stem Housing Vanguard CR Stem Housing Vanguard CR Stem Housing Vanguard CR Stem Housing Vanguard CR Stem Housing Vanguard CR Stem Housing 55 Left 57 5 Left 60 Left 62 5 Left 65 Left 67 5 Left 70 Left 72 5 Left 75...

Страница 8: ...er Guide 32 483077 32 483078 32 483079 32 483080 32 483081 32 483082 32 483083 32 483084 Vanguard Stem Housing Reamer Guide Bushing Vanguard Stem Housing Reamer Guide Bushing Vanguard Stem Housing Reamer Guide Bushing Vanguard Stem Housing Reamer Guide Bushing Vanguard Stem Housing Reamer Guide Bushing Vanguard Stem Housing Reamer Guide Bushing Vanguard Stem Housing Reamer Guide Bushing Vanguard S...

Страница 9: ...m Knee system c Critical factors in selecting the patients 1 Occupation or activity if the patient s occupation or activity involves making efforts with the knee the resulting force could cause a failure in the fixing and or implanting The implant will not restore function to the level of a healthy bone and patients should not be given false expectations 2 Senility mental illness or alcoholism the...

Страница 10: ...ies unless otherwise indicated This material is intended for the sole use and benefit of the Biomet sales force and physicians It is not to be redistributed duplicated or disclosed without the express written consent of Biomet For product information including indications contraindications warnings precautions and potential adverse effects see the package insert herein and Biomet s website ...