SYSTEM OVERVIEW
HOW TO TURN ON VIDAS?
Switch on the UPS then VIDAS
®
analyzer with the power switch locate on the back
of the analytical module.
Document only for Training use. Does not replace the User Manual.
Stand-alone pilot PC
External sample barcode reader
Retractable strip preparation tray
under each reagent strip tray
Allow the VIDAS
®
analyzer to warm up for 45 minutes.
Switch on the printer, the screen and the computer.
Wait for the Windows
®
welcome window to appear and enter the user’s name and
password.
After 2 or 3 minutes waiting, double click the VIDAS
®
PC icon.
After the 45 minutes warm-up phase, turn the VIDAS
®
analyzer power switch to OFF.
Wait 1 minute and turn it ON again.
This operation enables the standard value to be memorized after the
temperature has stabilized.
Trick: barcode wand must be plugged when system is OFF.
Analytical module: 5 independent sections of 6 slots
Содержание VIDAS