SYSTEM OVERVIEW
SPR
®
loading
position
Reagent loading position
(12 positions)
Integrated PC
Integrated Printer
Section
Sample Extrenal
Barcodes reader
HOW TO TURN ON MINI VIDAS?
Switch on the MINI VIDAS
®
with the power switch located on the back of
the analytical module.
Allow the MINI VIDAS
®
to warm up for
45 minutes
.
During initialization, the optical system automatically calibrates against a standard
which reacts to temperature.
Switch off the MINI VIDAS
®
.
Wait 1 minute and switch on the system again.
This operation enables the standard value to be memorized after the
temperature has stabilized.
Trick: barcode wand must be plugged when system is OFF.
MENU OVERVIEW
MAIN MENU
Start Section:
To start an analysis as soon as SPR
®
, strips and samples are in position.
Status Screen:
To know the status of sections A & B and temperature of SPR
®
and Strip areas.
Master Lot Menu:
To scan the Master Lot data for any new parameter or new assay lot.
Results Menu:
Direct access to stored results for validation, print or reprint of assay list.
Utility Menu:
Confi guration of the instrument (e.g.: unit, date & time, user ID, language,
printer, barcode wand).
TOUCH PAD: KEY FUNCTIONS
To erase a data
entry
To go back to
the previous screen
Press once: it will
display errors
or
Press twice: to print
out the screen
To move the cursor
on certain screens
To show a 2
nd
3
rd
screen
The «ENTER» key to
validate data
Document only for Training use. Does not replace the User Manual.
Refer to MINI VIDAS
®
user’s manual for detailed data