8. Technical specification
8.1. SIAscope V
•
Function - Non-invasive skin imaging device displaying the distribution and relative concentration
of blood, pigment (such as melanin), and collagen within the papillary dermis and/or epidermis.
Data is displayed as colour bitmaps known as SIAscans.
•
Sensing technology - Spectrophotometric Intracutaneous Analysis
•
Sensor - Hand held sensor with a 11mm diameter sensing area, linked by a standard USB 2.0
cable to the system unit.
•
Emitted wavelengths - 440nm to 960nm
•
Resolution - Features larger than 25 microns
•
Scan time - Approximately 3 seconds data collection, 3 seconds data processing
•
Data display format - Processed data displayed as colour bitmaps (SIAscans) representing relat-
ive concentration and distribution of selected chromaphores.
•
Reference image format - Enhanced colour ELM (dermatoscopic) image. Image calibrated for re-
peatability and optimized for maximum colour discrimination in both light and dark areas of the im-
age.
•
Calibration - Factory Calibrated
•
Dimensions - 240mm in length x 65mm in diameter at its widest point.
•
Weight - 250g
•
Viewfinder - Colour viewfinder mode allows accurate positioning of the sensor on the skin.
•
Hygiene - Easily disinfected sensor window. Wipe clean with an Isopropyl wipe before each scan.
•
Power supply - Power provided via the standard USB 2.0 connection.