28
Fault
Code
Cause
Effect
Actions
Error codes
555 ERROR_TREATMENT_PARAMETER_NROFREPETITIONS
Treatment
corrupted
Contact Biolight
service centre
556 ERROR_TREATMENT_PARAMETER_IMAGEOUTPUT
Treatment
corrupted
Contact Biolight
service centre
601 ERROR_THERMISTOR_TEMP_HIGH
Treatment is
shut down.
Can’t run
treatment
Wait to cool
down BCD
602 ERROR_LED_DRIVER_TEMP_HIGH
Treatment is
shut down.
Can’t run
treatment
Wait to cool
down BCD
603 ERROR_UNKNOWN_TREATMENT_TYPE
Failed to
identify
treatment
types from
credit
Contact Biolight
service centre
604 ERROR_UNKNOWN_INSTRUMENT_TYPE
Failed to
identify BCD
type from
credit
Contact Biolight
service centre
701 ERROR_CORRUPT_CREDIT_FILE
Credits not
stored to BCD
Contact Biolight
service centre for
a new credit file
702 ERROR_OLD_CREDIT_FILE
Credits not
stored to BCD
This credit file
is already
loaded to the
BCD
Use a new credit
file
900 ERROR_STARTUP_MODE_UNDEFINED
Reconnect
power cable
If code reoccurs
contact Biolight
service centre
901 ERROR_STARTUP_MODE_OVERLOAD
Remove power
cable
Check that
correct power
supply is used
303181_M17001_ver0_Biolight_Equine_Manual_Animal_ENG_A5.indd 28