ULT25NEU-LIT-OM-E
|
stirlingultracold.com
|
17 OF 34
To change the Setpoint by
Method B
>
Press and hold the
Set Temp
button until the
Set Temp LED
glows.
–
The Setpoint temperature will appear with the letter S in the first
LCD display.
>
The
^
/
v
buttons can be used to change the Setpoint.
–
When the
^
/
v
buttons are pushed the Setpoint will change by 1ºC at first.
–
If the
^
/
v
buttons are held for five seconds the step size will increase
to 5ºC.
–
Briefly release (a half second) the
^
/
v
buttons while continuing to
hold down the
Set Temp
button the increment/decrement size will
return to 1ºC.
–
Release the
^
/
v
buttons once the desired Setpoint is achieved.
>
Press the
Set Temp
button briefly (less than one second) to turn off
the
Set Temp LED
. The display will return to chamber temperature.
4.3
USING PRESET TEMPERATURES
The Temperature Presets — 1, 2 and 3 can be used to quickly change the
Setpoint to frequently used values. When shipped the Presets are set to
-86ºC, -40ºC, and -20ºC.
To choose one of the preset temperatures:
>
Press and hold the
Set Temp
button until the
Set Temp LED
glows.
>
Push one of the Temperature Preset buttons briefly.
>
The new Set Temperature will be shown in the LCD display.
>
If this is the desired Set Temperature, the user can either:
–
Briefly press the Set Temp button to turn off the
Set Temp LED
.
OR
–
Wait for approximately eight seconds when the freezer control panel
will return to normal operation.
4.
OPERATION
(continued)
Содержание Stirling Ultracold ULT25NEU
Страница 1: ...UL T25NEU Operating Manual 86 C Ultra Low Temperature Freezer...
Страница 4: ...4 OF 34 OPERATING MANUAL ULT25NEU 10 05 22 TABLE OF CONTENTS THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 32: ...32 OF 34 OPERATING MANUAL ULT25NEU 10 05 22 TABLE OF CONTENTS THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 33: ...ULT25NEU LIT OM E stirlingultracold com 33 OF 34 TABLE OF CONTENTS THIS PAGE INTENTIONALLY LEFT BLANK...