WARRANTY

We grant you a year warranty after the date of purchase. Any damage causes 
by improper handling shall not be covered by the warranty. Batteries and 
packaging are also excluded from the warranty. All other damage claims excluded. 
A warranty claim must be submitted with the purchase receipt.

V1.0

                                                                  

TECHNICAL SPECIFICATION

Type                                                      Maximum thermometer

Measuring range                                 32.0 C~ 42.9 C

                                                              Temp.<32.0 C: display  Lo for low

"

"

                                                              Temp.>42.9 C: display  Hi for high

" "

Resolution                                           0.1 C

Accuracy                                              

+

 0.1 C for measuring range 35.5 C ~ 42.0 C 

                                                              

0.2 C for other measuring range

Response time(in water bath)           23

+

2 C     37

+

2 C, within 15 seconds

Measurement time                              Oral: 20

+

5 sec. Underarm: 30

+

5 sec.

                                                              (for reference only, it differs from people to people)

Power source                                      One 1.5V 

 

 battery, type LR41, 192, SR41, 392 or equivalent 

Power consumption                           0.15mW(Max.)in the measuring mode

Memory function                                 Turn on the unit it automatically display the last reading for 2

                                                              seconds after display check.

Auto Power-Off function                    Around 8 min. 40sec. 

+

1min. after the measurement completed

Waterproof function                           The whole unit is resistant to water

Operating temperature                     10~40 C, 15~95%R.H.  noncondension

Storage temperature                          -20 C to + 60 C

Self-test function                                It runs automatically after power on. Display Err when system 

"

"

                                                              malfunction

Reference to standards                     EN12470-3:2000,clinical thermometers

                                                              ASTM E1112-00 standard

Signaling tone                                     For signaling that the thermometer is ready to use and that the

measurement is completed

 Rectal: 10+5 sec.

CLEANING AND DISINFECTING

Note:

The thermometer is 100% waterproof. You can clean it by washing with soap and 
warm water or by wiping it with  70% isopropyl alcohol. Then wipe dry with a soft, 
clean cloth.
DO NOT BOIL OR CLEAN IN THE DISHWASHER. This will damage your 
thermometer and will void the warranty.

 Any other cleaning solution, dilution or cleaning method can cause a 

malfunction of the thermometer!