Second Edition, Version 2.03
April 2017
PlasmaDive
™ Mini Manual, © Biognosys AG, Switzerland
Page 21 of 22
Issue
Possible Cause
Recommended Solution
Only iRT peptides are
displayed in
SpectroDive
™ analysis
tree
The corresponding
Panel Plug-in is not
correct or not installed
Install the correct Panel
Plug-in for the kit you are
using.
Analytical run was
performed with
incomplete transition list
(e.g. without transitions
for panel peptides)
Check if the transition list
contains panel peptides. If
not, do the calibration
according to the protocol
(Section F) and re-analyze
the samples with the
corrected MS-method.
Exported report is empty
or does not correspond
to the description in the
manual
A correct SpectroDive
™
Panel Plug-in is missing
Install a correct Panel Plug-in
for the kit you are using.
The incorrect report
schema was chosen in
step 101 when
generating the report
Choose the correct report
schema and export the
report again.
Some of the protein
quantities in my report
are shown as "Not
Detectable"
Low quality of the
analytical run
(unsatisfactory LC-MS
performance)
Control performance of your
LC-MS system and repeat
the sample analysis.
The protein
concentrations are
below the limit of
quantification
Detectability of target
proteins may vary between
samples as some of the
proteins are only detectable
in certain medical conditions.