Gene Q Thermal Cycler
Chapter 4 Operation Guide
—
15
—
(Please refer to the following figure: relation between block temperature and reagent
temperature.)
“Tube mode”
has considered the phenomenon that the cooling & heating rate is a little
slower than that of block. There is an overshoot after reaching the setting temperature, which
can allow reagent to reach the setting temperature faster. During the same setting time as
“Block”, the actual time for reagent to keep the setting temperature is prolonged obviously.
(Please refer to the following figure: relation between block temperature and reagent
temperature.)
3.5
“Hotlid” stand for the state of hotlid “Open” or “Close”.
3.6
“Key Sound” and “Run End Sound” stand for the sound of pressing key and reminding sound
after the procedure runs over respectively.
Block
Reagent
Block
Reagent