Operation Manual for Automatic Nucleic Acid Purification System

I

Important Notes

1 General Practice

Caution

:

Because of the important information in this column, please read it carefully. Failure to

follow the advice in this column will possibly result in damage or malfunctioning of

the machine.

2 Safety

During the operation, maintenance, or repair of the machine, the following safety measures should be
taken. Otherwise, the guard provided by the machine is likely to be damaged, the rated safety level to
be reduced, and the rated operation conditions to be affected.
Hangzhou Bioer Technology Co. Ltd. shall not be in any way responsible for the consequences
resulted from buyer’s not observing the following requirements.

a) Grounding Considerations

A.C. power’s grounding should be reliable for fear of an electric shock. The 3-pin plug with the
machine’s power line is a safety device that should be matched with a grounded socket. Never let the
third ground pin floating. If the 3-pin plug cannot be inserted, it is recommended to ask an electrician
to install an appropriate power socket.

b) Keep Away from Electric Circuits

The operator is not allowed to open the machine. Changing components or adjusting certain
parameters inside the device must only be accomplished by the certificated professional maintenance
personnel. Do not change elements while the power is still on.

c) A.C. Power Considerations

Before turn on the power, always check and insure that the mains voltage is within the required A.C.
100-240V and insure that the current rating load of the power socket meet the required specification.

d) A.C. Power Line Considerations

As an accessory of the machine, the A.C. power line should be a default one. If it is damaged, the A.C.
power line can not be repaired, but must be replaced with a new one. The power line should be free of
heavy objects during the machine’s operation. Keep the power line away from the place where people
ambulate frequently.

e) Connect the A.C. Power Line

While connect or disconnect the power line, you should hold the 3-pin plug with your hand. Insert the
plug thoroughly into the socket to ensure good contact between the plug and socket. Pull the plug, not
the line, when you need to disconnect to the mains.

Содержание GenePure Pro NPA-32P

Страница 1: ...GenePure Pro Nucleic Acid Purification System NPA 32P Users Manual Hangzhou Bioer Technology Co Ltd...

Страница 2: ...ithout prior written consent from Hangzhou Bioer Technology Co Ltd any part of the manual shall not be duplicated reproduced or translated into any other languages Product Technical Requirements No Z...

Страница 3: ...mmended to ask an electrician to install an appropriate power socket b Keep Away from Electric Circuits The operator is not allowed to open the machine Changing components or adjusting certain paramet...

Страница 4: ...f the following events occurs you are suggested to disconnect the power line with mains and contact the distributor or ask a certificated maintenance worker for help Liquid into the device The device...

Страница 5: ...ds please immediately contact the supplier After qualification of acceptance properly fill in the product acceptance sheet and send or fax the copied sheet to the supplier for filing and maintenance A...

Страница 6: ...rtation and Storage Conditions 2 3 Basic Parameters 2 4 Basic Performance 3 5 Software Functions 3 Chapter III Preparations 4 1 Structure Diagram 4 2 Check before Startup 5 3 Placement of 96 deep well...

Страница 7: ...follows small volume light weight low noise full closed workaround Embedded real time operating system Transparent working chamber facilitating the observation of working conditions Large LCD panel ea...

Страница 8: ...out condensation Power supply AC100 240V 50 60Hz Caution Before power on please confirm whether the working conditions meet the above requirements Pay special attention to the power line is reliably g...

Страница 9: ...am storage 20 5 Software functions File edit view modify and delete functions File run pause and stop functions File import export by U disk Voice prompt function Fault protection and alarm function U...

Страница 10: ...on the screen as well as the preparations before the startup Before the first operation of this instrument please read this chapter carefully and make sure you are familiar with the contents of this c...

Страница 11: ...o help fix the 96 deep well plate b Note Make sure the 96 deep well plate is placed evenly before starting up 4 Startup Switch on the automatic nucleic acid purification system and the instrument will...

Страница 12: ...off the power immediately and contact Bioer Ready Let s get started 1 Upper computer operating interface a Operation main interface After the system is started enter the main interface of nucleic acid...

Страница 13: ...u items the bottom function button and browsing information will be different The following figure shows the file browsing interface A list of stored files is in the middle showing up to 9 files per p...

Страница 14: ...be displayed when the U disk is inserted If the U disk is not inserted it prompts failure to detect the U disk In the center is the ebr file list that conforms to the file name standard in the default...

Страница 15: ...uding well location step name waiting time mixing time binding time volume mixing mode and binding mode Below steps interface are function buttons including Add Delete Temperature Previous Page Next P...

Страница 16: ...ng interface as shown in the following figure Set up well positions 1 6 liquid volume waiting time mixing mode and binding mode Click on the list box beside the name and the name selection list will p...

Страница 17: ...set the time to 00 00 The mixing speed can be selected from the option list by clicking on the mixing steps list See the figure below Click on the list box beside the binding mode and the binding par...

Страница 18: ...wsing interface Click on Run button in the history browsing interface Click on Run button in the file save interface The lysis and elution temperatures are displayed on top of the running interface Th...

Страница 19: ...click cancel button to return to the main interface f System setting Click on the system settings to enter the system settings interface in the operation main interface The 8 buttons in the middle co...

Страница 20: ...ettings interface i Backlight adjustment Click backlight adjustment on the system setting interface to enter the backlight adjustment interface Turn up down screen brightness by clicking and dragging...

Страница 21: ...turn to the system setting interface l Screen calibration In the operation main interface click the screen calibration to enter the screen calibration interface and click the center of the cross icon...

Страница 22: ...ll be indicated that the upgrade file is not detected After selecting the program that needs to be upgraded according to the upgrade file in the USB flash disk automatically list the items available f...

Страница 23: ...anufacturer 2 When power switch turned on buzzing alarm sounds and system error is displayed Please contact the manufacturer Error occurs during self inspection process Contact the supplier or manufac...

Страница 24: ...Bin An Rd Hi Tech Binjiang District Hangzhou City 310053 China Tel 0571 85800535 87774558 Fax 0571 85800537 87774559 E mail overseas bioer com cn www bioer com cn Han gz h o u Bi oe r Te c h n ol o g...

Отзывы: