7. Adjust height of assembly on stand for patient comfort. Explain 

procedure to the patient.  Place disposable gloves on the patient if 
handling of the tubing will be required by the patient during the study.

8. Place the mouthpiece in the patient’s mouth. After ensuring the patient 

maintains a good seal on the mouthpiece, apply the nose clip and 
instruct the patient to relax and breathe deeply on the system several 
times to ensure proper operation and familiarize the patient with the 
system.

9. Connect one end of O

2

tubing to the O

2

supply and set the flow to 

10-11 liters per minute. Confirm that the patient is ready to begin the 
procedure. Connect the other end of the O

2

tubing to the Venti-Scan IV 

inlet tube to start the aerosol delivery procedure. INSTRUCT THE 
PATIENT TO CONTINUE BREATHING ON THE SYSTEM FOR THREE 
TO FIVE MINUTES TO ENSURE SUFFICIENT LUNG DEPOSITION.

10. After allowing enough time to ensure sufficient lung deposition, turn 

OFF the oxygen (or air). To maximize clearance of any remaining 
activity in the tubing following inhalation, and to minimize the possibility 
of inadvertent contamination to the patient and room, instruct the 
patient to continue breathing for five or six breaths before removing the 
mouthpiece (you may also instruct the patient to expectorate into a 
disposable towel to be handled as normal radioactive waste).

11. Once the patient mouthpiece has been removed, remove the nose clip 

and begin imaging.

NOTE:

Imaging may be performed during radioaerosol inhalation.

CAUTION: 

To prevent inadvertent contamination of the patient or

room, use care and good radiation practice when removing the patient
from the system.

ATTENTION: 

Pour éviter de contaminer le patient et la piece, suivre

les pratiques usuelles relatives aux rayons X lors de la sortie du patient
du système.

CAUTION: 

This procedure may NOT be appropriate for uncooperative

patients or for patients who cannot maintain a good mouthpiece seal.

ATTENTION:

Cette technique peut ne pas être adaptée à des patients

non coopérants ou à des patients ne pouvant pas maintenir une bonne
étanchéité sur la pièce buccale.

12. Disconnect the oxygen tube. Remove the shield assembly from the IV 

pole mount and transport to a radioactive waste disposal area.

Содержание Venti-Scan IV

Страница 1: ...iV RadioaeRoSol conVenience Kit operation manual 177 091 177 092 BIODEX Biodex Medical Systems Inc 20 Ramsey Road Shirley New Y ork 11967 4704 Tel 800 224 6339 Int l 631 924 9000 Fax 631 924 9241 Emai...

Страница 2: ...om top to bottom inStRuCtiOnS FOR uSe 1 Remove the Venti Scan IV shielded canister top 2 Place lower shield assembly in IV pole mount 3 Remove the Venti Scan IV l kit from its packaging 4 Gently inser...

Страница 3: ...vity in the tubing following inhalation and to minimize the possibility of inadvertent contamination to the patient and room instruct the patient to continue breathing for five or six breaths before r...

Страница 4: ...and moving it to the decay barrel The used circuit contains significant radioactive material Be certain to use your facility approved procedure for disposal of all radioactive and biohazardous waste N...

Страница 5: ...deep breaths Patients with false teeth or no teeth or a large amount of hair in the nose need to be monitored closely They may move the mouthpiece around and have mist squirt out sides of mouth or the...

Страница 6: ...45 psi of pressure When the tubing is connected and used by the patient the pressure drops to about 28 psi This is still ample to get small partial vents 1 to 1 2 micron mass mean aerosolized diameter...

Страница 7: ...you have your mismatch or possible PE If not the patient may have COPD NOTE The camera must not have a cooling fan near the crystal head If the patient coughs talks or spits out the mouthpiece you cou...

Страница 8: ...BIODEX Biodex Medical Systems Inc 20 Ramsey Road Shirley New Y ork 11967 4704 Tel 800 224 6339 Int l 631 924 9000 Fax 631 924 9241 Email info biodex com www biodex com MADE IN U S A...

Отзывы: