![biodex ATOMLAB 100 Скачать руководство пользователя страница 2](http://html1.mh-extra.com/html/biodex/atomlab-100/atomlab-100_manual_2753512002.webp)
Setting Isotope Selection Keys, 1, 2 and Labeled Isotope Keys
1.
Press appropriate ISOTOPE SELECTION KEY
2.
Press SETUP
3.
The calibration value for selected isotope is displayed on ETD (#1)
4.
Use the
⇑⇓
keys to set new calibration value on ETD (#1)
5.
Press ENTER
6.
Press SETUP
Setting OTHER Key Isotope Values
1.
Press OTHER KEY
2.
Press SETUP
3.
Use
⇑⇓
keys to set new calibration value on ETD (#1)
4.
Press ENTER
5.
Press SETUP
6.
System returns to normal operating mode – the new calibration value is stored
Isotope default Calibration Values
1.
Press any standard ISOTOPE SELECTION KEY
2.
Press SETUP
3.
Current default value for selected isotope is displayed on ETD (#1).
Compare this value to “Appendix D” in Operation Manual
4.
Use the
⇑⇓
keys to adjust isotope default value as needed on ETD (#1)
5.
Press ENTER
6.
Press SETUP
7.
System returns to normal operating mode – the new calibration values are stored
The following method can also be used to determine the dial setting for a calibrated standard.
OTHER Key Activity Measurement
1.
Press OTHER KEY
2.
ETD (#1) displays dial setting for OTHER key
3.
Use the
⇑⇓
keys to adjust isotope default value as needed on ETD (#1)
4.
Put source in well and take reading
5.
Press another isotope key
6.
Press OTHER and dial setting is back to default setting
B
IODEX
Biodex Medical Systems, Inc.
20 Ramsey Road, Shirley, New York, 11967-4704, Tel: 800-224-6339 (
Int’l 631-924-9000),
Fax: 631-924-9241, Email: [email protected], www.biodex.com
M A D E I N
U . S . A .
+ + + + + +
+ + + + +
+ + + + + +
+ + + + +
+ + + + + +
+ + + + +
+ + + + + +
+ + + + +
+ + + + + +
Certified Quality Management System