Blood Pressure Owner’s Manual
Blood Pressure Owner’s Manual
Blood Pressure Owner’s Manual
Blood Pressure Owner’s Manual
Blood Pressure Owner’s Manual
55
15
b. Touch START button, pump begins to inflate the cuff. The rising
pressure in the cuff is shown on the display.
NOTE
In case the inflation pressure is not sufficient to determine measurement
result, the monitor automatically re-inflates to a higher level.
c. Pressure in cuff increases to working pressure. Then pump stops
and pressure falls.
d. The heart symbol “ ” starts to flash every heart beat once pulse is
detected.
e. fter the measurement, systolic pressure, diastolic pressure, pulse rate
and color bars classified from ESH (European Society of Hypertension)
standards shows on the display.
f. If irregular heartbeat was detected during the measurement, LCD
displays the “ ” icon to remind users of heartbeat irregularity. Please
consult your doctor if it appears often.
CAM MODE
(CLINICALLY ACCURATE MEASUREMENT)
Always apply the pressure cuff before turning on the monitor.
In CAM mode, 2 measurements (at 20 sec break) are automatically taken
in succession and the result is then automatically analyzed and displayed.
Because people’s blood pressure constantly fluctuates, a result
determined in this way is more reliable than the one produced by a single
measurement.
a. Turn on the device by pressing the O/I button at the back side of the
monitor.
b. Touch CAM button, pump begins to inflate the cuff. LCD display shows
CAM icon and the rising pressure in the cuff.
c. Pressure in cuff increases to working pressure. Then pump stops and
pressure falls.
NOTE
In case the inflation pressure is not sufficient to determine measurement
result, the monitor automatically re-inflates to a higher level.
d. The heart symbol “ ” starts to flash every heart beat once pulse is
detected.
e. After the 1st measurement, the monitor automatically starts the 2nd
measurement after 20 sec countdown.
f. After the 2nd measurement, the device analyzes the1st and 2nd
measurement values: if sys BP1- sys BP 2 >/= 15 mmHg, dia BP1 - dia BP2
>/= 10 mmHg, CAM mode will automatically start the 3rd measurement
after another 20 sec countdown.
NOTE
In case the 1st and/or 2nd pulse rate </= 60, device willautomatically take
4 times measurement in succession.
g. The CAM measured systolic pressure, diastolic pressure, pulse rate and
color bars classified from ESH( European Society of Hypertension)
standards show on the display.
h. If irregular heartbeat was detected during the measurement, LCD
displays the “ ” icon to remind users of heartbeat irregularity.
Please consult your doctor if it appears often.
i. If LCD displays the “ ” icon quite often, please consult your doctor
to check what kind of arrhythmias you may have. It is important for
early diagnostic of atrial fibrillation.
MONITOR MEMORY
Press O/I button at the back side of the device, the monitor is turned
on and stays standby.
AVERAGE OF ALL STORED MEMORIES
Touch M/SET button, LCD shows a set of values averaged from all the
stored memories of the selected user, accompanied by an icon “A”.
AVERAGE OF ALL STORED DAY-TIME MEMORIES
Press M/SET button again, LCD shows another set of values averaged
from all the stored day-time (4:00~9:59) taken measurements,
accompanied by an icon “A” and .
NOTE
Heart attack and stoke occurs more frequently in the morning. This
feature averages the morning readings between 4:00 and 9:59AM to
ident i t y morning hypertensions.
AVERAGE OF ALL STORED NIGHT-TIME MEMORIES
Press M/SET button once more, LCD shows the other set of values
averaged from all the stored night-time (17:00~1:59) taken
measurements, accompanied by an icon “A” and .
MEMORY RECALL
The device stores 90 most recent blood pressure test results along with
respective dates and times in the monitor memory for each of two users.
Reviewing Test Results
a. Touch M/SET button again after you see the average of
all stored night-time memories, the single memory
reading appears.
b. Per the setting, the first reading you see is the last blood
pressure result. Touch M/SET button again, the next
measurement eg. M89, as well as the previous 88
measurements (M88, M87… M1) can be displayed
one after the other.
NOTE : CAM icon will show on LCD if the measurement is taken under CAM mode.
DELETE ALL STORED MEMORIES.
With device in standby status, hold down M/SET button for 14 seconds
until “CL” flashes on LCD, press M/SET button again, all the stored
memories inside is removed.
Before you delete all the readings stored in the memory, please make
sure you will not need to use them any more at a later time.
DOWNLOADING RESULTS ONTO A COMPUTER
This device is able to be used with a personal computer running URIGHT
Health Care Management System. All stored memory readings can be
transferred to your computer via the attached micro USB cable.
a. Software installation Download the software from website, then click
setup for installation of BioCare corproation.
b. Data Transmission Connect the monitor to your computer via micro
USB cable, LCD immediately displays three lines to indicate successful
connection and data transmission is now available.
NOTE
During the whole process of USB connection, the monitor is completely controlled by the
software system.
MAINTENANCE
CARING FOR YOUR MONITOR
To avoid the monitor attracting dirt, dust or other contaminants, wash and dry your hands
thoroughly before use.
Cleaning
• To clean the monitor exterior, wipe it with a cloth moistened with tap water or a mild
cleaning agent, then dry the device with a soft dry cloth. Do NOT flush with water.
• Do NOT use organic solvents to clean the monitor.
• Do NOT wash the pressure cuff.
• Do NOT iron the pressure cuff.
Monitor Storage
• Storage condition: -20°C to 55°C (-4°F to 131°F), below 95% relative humidity.
• Always store or transport the monitor in its original storage case.
• Avoid dropping or heavy impact.
• Avoid direct sunlight and high humidity.
DETAILED INFORMATION
Reference Values
Clinical studies show that adult diabetes is often accompanied by elevated blood pressure.
People with diabetes can reduce their heart risk by managing their blood pressure along
with diabetes treatment*
4
.
Monitoring your routine blood pressure trend helps you to know your body condition.
Human blood pressure naturally increases after reaching middle age. This symptom is a
result of continuous ageing of the blood vessels. Further causes include obesity, lack of
exercise and cholesterol (LDL) adhering to the blood vessels. Rising blood pressure
accelerates hardening of the arteries, and the body becomes more susceptible to apoplexy
and coronary infarction. The recommended blood pressure range is as below :
Isolated systolic hypertension should be graded (1, 2, 3) according to systolic blood pressure values in the ranges indicated, provided
that diastolic values are < 90mmHg.
Source: The European Society of Hypertension and European Society of Cardiology Task Force Members. 2007 ESH-ESC Practice
Guidelines for the Management of Arterial Hypertension. J Hypertens 2007; 25: 1751-1762.
*1 American Diabetes Association. The Diabetes-Heart Disease Link Surveying Altitudes Knowledge and Risk (2002)
SYSTEM TROUBLESHOOTING
If you follow the recommended action but the problem persists, or error messages other
than the ones below appear, please call your local customer service. Do not attempt to
repair by yourself and never try to disassemble the monitor under any circumstances.
Error Messages
Troubleshooting
1. If no display appears after pressing Power button
2. If the heart rate is higher/lower than user’s average
3. If the result is higher/lower than user’s average measurement
4. If the cuff inflates again while measuring
SYMBOL INFORMATION
SPECIFICATIONS
System Performance
Blood pressure measurement performance
This device has been tested to meet the electrical and safety requirements of :
IEC/EN 60601-1,
IEC/EN 60601-1-2
Reference to Standards :
• EN 1060-1 /-3, NIBP-requirements
• IEC60601-1 General requirement for safety
• IEC60601-1-2 Requirements for EMC
• EN1060-4, NIBP clinical investigation
• AAMI / ANSI / IEC 80601-2-30, ANSI/AAMI/ISO 81060-2, NIBP requirements
Category
Optimal
Normal*
3
High normal
Grade 1 hypertension
Grade 2 hypertension
Grade 3 hypertension
Isolated systolic hypertension
Systolic (mmHg)
< 120
120–129
130–139
140 – 159
160–179
≥ 180
≥ 140
and
and/or
and/or
and/or
and/or
and/or
and
Diastolic (mmHg)
< 80
80–84
85–89
90 – 99
100–109
≥ 110
< 90
MESSAGE CAUSE WHAT TO DO
Inflation or pressure
error
Blood pressure
measurement error
Refit cuff tightly and correctly.
If error still remains, contact local
customer service for help.
Refit cuff tightly and correctly.
Relax and repeat the measurement.
If error still remains, contact local
customer service for help.
Relax and repeat the measurement.
Improper movement
detected during
measurement (Artifact).
No pulse detected
Problems with the
monitor
Refit cuff tightly and correctly.
Relax and repeat the measurement.
If error still remains, contact local
customer service for help.
Review the instructions and repeat
the test.
If the monitor still does not work,
please contact the local customer
service for help.
POSSIBLE CAUSE
Batteries exhausted.
Batteries incorrectly installed or absent.
WHAT TO DO
Replace the batteries.
Check that the batteries are correctly
installed.
POSSIBLE CAUSE
Movement during measurement.
Measurement taken just after exercise.
WHAT TO DO
Repeat measurement.
Rest at least 30 minutes beforehand.
POSSIBLE CAUSE
May not be in correct position while
measuring.
Blood pressure naturally varies from
time to time.
WHAT TO DO
Adjust to the correct position to
measure.
Keep in mind for next measurement.
POSSIBLE CAUSE
Cuff is not fastened.
If user’s blood pressure is higher than the pressure the device has inflated, the
device will automatically increase the pressure and start to inflate again. Stay
relaxed and wait for the measurement.
WHAT TO DO
Fasten the cuff again.
Blood Pressure Owner’s Manual
Symbol Referent Symbol Referent
Consult instructions for use
Authorized representative in
the European Community
Manufacturer
Class II
Humidity limitation
Water-proof grade
CE mark
Serial number
Caution, consult
accompanying documents
Type BF Equipment
Dispose of in accordance with
the requirements of your
country.
Power source
Size of monitor w/o cuff
Weight
Cuff Size
Memory
Power saving
System operating condition
Meter storage condition
Pressure range
Heart rate range
Systolic Measurement Range
Diastolic Measurement Range
Pulse Rate Measurement Range
Maximum inflation pressure
Accuracy of Pressure
Accuracy of Pulse rate
Measurement unit
Four 1.5V AAA alkaline batteries,5V 0.6A mains adaptor
138 (L) x 94 (W) x 33 (H) mm,
260g ± 5g with batteries
M (medium) : 22-38cm / 8.7”~14.9”
Maximum 90 memory records
Automatic power off if system idle 1.5 min.
10
o
C to 40
o
C (50
o
F to 104
o
F),
15-95% relative maximum humidity
-20
o
C to 55
o
C (-4
o
F to 131
o
F),
15-95% relative maximum humidity
20 - 280 mmHg
40 - 200 beat per minute
50 mmHg -250 mmHg
30 mmHg -180 mmHg
40 - 200 beats / minute
280 mmHg
±3 mmHg or ±2% of the reading
±4% of the reading
mmHg