Blood Pressure Owner’s Manual

Blood Pressure Owner’s Manual

Blood Pressure Owner’s Manual

Blood Pressure Owner’s Manual

Blood Pressure Owner’s Manual

55

15

b. Touch START button, pump begins to inflate the cuff. The rising 

     pressure in the cuff is shown on the display.

NOTE

In case the inflation pressure is not sufficient to determine measurement

result, the monitor automatically re-inflates to a higher level.

c. Pressure in cuff increases to working pressure. Then pump stops

    and pressure falls.

d. The heart symbol “        ” starts to flash every heart beat once pulse is 

    detected.

e. fter the measurement, systolic pressure, diastolic pressure, pulse rate 

    and color bars classified from ESH (European Society of Hypertension)

    standards shows on the display.

f. If irregular heartbeat was detected during the measurement, LCD 

    displays the “         ” icon to remind users of heartbeat irregularity. Please 

    consult your doctor if it appears often.

CAM MODE

(CLINICALLY ACCURATE MEASUREMENT)

Always apply the pressure cuff before turning on the monitor.

In CAM mode, 2 measurements (at 20 sec break) are automatically taken 

in succession and the result is then automatically analyzed and displayed.

Because people’s blood pressure constantly fluctuates, a result 

determined in this way is more reliable than the one produced by a single

measurement.

a. Turn on the device by pressing the O/I button at the back side of the 

    monitor.

b. Touch CAM button, pump begins to inflate the cuff. LCD display shows 

    CAM icon and the rising pressure in the cuff.

c. Pressure in cuff increases to working pressure. Then pump stops and 

    pressure falls.

NOTE

In case the inflation pressure is not sufficient to determine measurement 

result, the monitor automatically re-inflates to a higher level.

d. The heart symbol “        ” starts to flash every heart beat once pulse is 

    detected.

e. After the 1st measurement, the monitor automatically starts the 2nd 

     measurement after 20 sec countdown.

f. After the 2nd measurement, the device analyzes the1st and 2nd 

   measurement values: if sys BP1- sys BP 2 >/= 15 mmHg, dia BP1 - dia BP2

    >/= 10 mmHg, CAM mode will automatically start the 3rd measurement 

   after another 20 sec countdown.

NOTE

In case the 1st and/or 2nd pulse rate </= 60, device willautomatically take

4 times measurement in succession.

g. The CAM measured systolic pressure, diastolic pressure, pulse rate and 

    color bars classified from ESH( European Society of Hypertension) 

     standards show on the display.

h. If irregular heartbeat was detected during the measurement, LCD

     displays the “            ” icon to remind users of heartbeat irregularity.

     Please consult your doctor if it appears often.

i. If LCD displays the “         ” icon quite often, please consult your doctor 

    to check what kind of arrhythmias you may have. It is important for 

    early diagnostic of atrial fibrillation.

MONITOR MEMORY

Press O/I button at the back side of the device, the monitor is turned 

on and stays standby.

AVERAGE OF ALL STORED MEMORIES

Touch M/SET button, LCD shows a set of values averaged from all the

stored memories of the selected user, accompanied by an icon “A”.

AVERAGE OF ALL STORED DAY-TIME MEMORIES

Press M/SET button again, LCD shows another set of values averaged 

from all the stored day-time (4:00~9:59) taken measurements, 

accompanied by an icon “A” and          .

NOTE

Heart attack and stoke occurs more frequently in the morning. This 

feature averages the morning readings between 4:00 and 9:59AM to

ident i t y morning hypertensions.

AVERAGE OF ALL STORED NIGHT-TIME MEMORIES

Press M/SET button once more, LCD shows the other set of values 

averaged from all the stored night-time (17:00~1:59) taken 

measurements, accompanied by an icon “A” and         .

MEMORY RECALL

The device stores 90 most recent blood pressure test results along with

respective dates and times in the monitor memory for each of two users.

Reviewing Test Results

a. Touch M/SET button again after you see the average of 

    all stored night-time memories, the single memory 

    reading appears.

b. Per the setting, the first reading you see is the last blood

     pressure result. Touch M/SET button again, the next 

     measurement eg. M89, as well as the previous  88

     measurements (M88, M87… M1) can be displayed 

     one after the other.

NOTE  : CAM icon will show on LCD if the measurement is  taken under CAM mode.

DELETE ALL STORED MEMORIES.

With device in standby status, hold down M/SET button for 14 seconds 

until “CL” flashes on LCD, press M/SET button again, all the stored 

memories inside is removed.

Before you delete all the readings stored in the memory, please make 

sure you will not need to use them any more at a later time.

DOWNLOADING RESULTS ONTO A COMPUTER

This device is able to be used with a personal computer running URIGHT 

Health Care Management System. All stored memory readings can be 

transferred to your computer via the attached micro USB cable.

a. Software installation Download the software from website, then click

    setup for installation of  BioCare corproation.

b. Data Transmission Connect the monitor to your computer via micro

    USB cable, LCD immediately displays three lines to indicate successful

    connection and data transmission is now available.

NOTE

During the whole process of USB connection, the monitor is completely controlled by the 

software system.

MAINTENANCE

CARING FOR YOUR MONITOR

To avoid the monitor attracting dirt, dust or other contaminants, wash and dry your hands 

thoroughly before use.

Cleaning

• To clean the monitor exterior, wipe it with a cloth moistened with tap water or a mild 

   cleaning agent, then dry the device with a soft dry cloth. Do NOT flush with water.

• Do NOT use organic solvents to clean the monitor.

• Do NOT wash the pressure cuff.

• Do NOT iron the pressure cuff.

Monitor Storage

• Storage condition: -20°C to 55°C (-4°F to 131°F), below 95% relative humidity.

• Always store or transport the monitor in its original storage case.

• Avoid dropping or heavy impact.

• Avoid direct sunlight and high humidity.

DETAILED INFORMATION

Reference Values

Clinical studies show that adult diabetes is often accompanied by elevated blood pressure. 

People with diabetes can reduce their heart risk by managing their blood pressure along 

with diabetes treatment*

4

.

Monitoring your routine blood pressure trend helps you to know your body condition. 

Human blood pressure naturally increases after reaching middle age. This symptom is a 

result of continuous ageing of the blood vessels. Further causes include obesity, lack of 

exercise and cholesterol (LDL) adhering to the blood vessels. Rising blood pressure 

accelerates hardening of the arteries, and the body becomes more susceptible to apoplexy 

and coronary infarction. The recommended blood pressure range is as below :

Isolated systolic hypertension should be graded (1, 2, 3) according to systolic blood pressure values in the ranges indicated, provided 

that diastolic values are < 90mmHg. 

Source: The European Society of Hypertension and European Society of Cardiology Task Force Members. 2007 ESH-ESC Practice 

Guidelines for the Management of Arterial Hypertension. J Hypertens 2007; 25: 1751-1762.

*1 American Diabetes Association. The Diabetes-Heart Disease Link Surveying Altitudes Knowledge and Risk (2002)

SYSTEM TROUBLESHOOTING

If you follow the recommended action but the problem persists, or error messages other 

than the ones below appear, please call your local customer service. Do not attempt to 

repair by yourself and never try to disassemble the monitor under any circumstances.

Error Messages

Troubleshooting

1. If no display appears after pressing Power button

2. If the heart rate is higher/lower than user’s average

3. If the result is higher/lower than user’s average measurement

4. If the cuff inflates again while measuring

SYMBOL INFORMATION 

SPECIFICATIONS

System Performance

 Blood pressure measurement performance

This device has been tested to meet the electrical and safety requirements of :

IEC/EN 60601-1,

IEC/EN 60601-1-2

Reference to Standards :

• EN 1060-1 /-3, NIBP-requirements

• IEC60601-1 General requirement for safety

• IEC60601-1-2 Requirements for EMC

• EN1060-4, NIBP clinical investigation

• AAMI / ANSI / IEC 80601-2-30, ANSI/AAMI/ISO 81060-2, NIBP requirements

Category

Optimal

Normal*

3

High normal

Grade 1 hypertension

Grade 2 hypertension

Grade 3 hypertension

Isolated systolic hypertension

Systolic (mmHg)

< 120

120–129

130–139

140 – 159

160–179

≥ 180

≥ 140

and

and/or

and/or

and/or

and/or

and/or

and

Diastolic (mmHg)

< 80

80–84

85–89

90 – 99

100–109

≥ 110

< 90

         MESSAGE                   CAUSE                                             WHAT TO DO      

Inflation or pressure

error

Blood pressure 

measurement error

Refit cuff tightly and correctly. 

If error still remains, contact local 

customer service for help. 

Refit cuff tightly and correctly. 

Relax and repeat the measurement. 

If error still remains, contact local 

customer service for help.

Relax and repeat the measurement.

Improper movement

detected during

measurement (Artifact).

No pulse detected

Problems with the

monitor

Refit cuff tightly and correctly. 

Relax and repeat the measurement.

If error still remains, contact local

customer service for help.

Review the instructions and repeat 

the test. 

If the monitor still does not work, 

please contact the local customer 

service for help.

POSSIBLE CAUSE 

Batteries exhausted. 
Batteries incorrectly installed or absent.

WHAT TO DO

Replace the batteries.
Check that the batteries are correctly

installed. 

POSSIBLE CAUSE 

Movement during measurement.
Measurement taken just after exercise.

WHAT TO DO

Repeat measurement.
Rest at least 30 minutes beforehand.

POSSIBLE CAUSE 

May not be in correct position while

measuring.
Blood pressure naturally varies from

time to time.

WHAT TO DO

Adjust to the correct position to

measure.
Keep in mind for next measurement.

POSSIBLE CAUSE 

Cuff is not fastened.
If user’s blood pressure is higher than the pressure the device has inflated, the 

device will automatically increase the pressure and start to inflate again. Stay

relaxed and wait for the measurement. 

WHAT TO DO

Fasten the cuff again.

Blood Pressure Owner’s Manual

      Symbol                       Referent                            Symbol                      Referent

Consult instructions for use 

Authorized representative in 
the European Community

Manufacturer 

Class II

Humidity limitation

Water-proof grade

CE mark

Serial number 

Caution, consult
accompanying documents 

Type BF Equipment

Dispose of in accordance with

the requirements of your 

country. 

Power source
Size of monitor w/o cuff
Weight
Cuff Size
Memory
Power saving
System operating condition

Meter storage condition

Pressure range
Heart rate range
Systolic Measurement Range
Diastolic Measurement Range
Pulse Rate Measurement Range
Maximum inflation pressure
Accuracy of Pressure
Accuracy of Pulse rate
Measurement unit

Four 1.5V AAA alkaline batteries,5V 0.6A mains adaptor
138 (L) x 94 (W) x 33 (H) mm,
260g ± 5g with batteries
M (medium) : 22-38cm / 8.7”~14.9”
Maximum 90 memory records 
Automatic power off if system idle 1.5 min.
10

o

C to 40

o

C (50

o

F to 104

o

F),

15-95% relative maximum humidity
-20

o

C to 55

o

C (-4

o

F to 131

o

F),

15-95% relative maximum humidity

20 - 280 mmHg
40 - 200 beat per minute
50 mmHg -250 mmHg
30 mmHg -180 mmHg
40 - 200 beats / minute
280 mmHg
±3 mmHg or ±2% of the reading
±4% of the reading
mmHg