4
Related Parts / Products
Retrieval Solutions
Biocare Medical retrieval solutions are specially formulated for pH stability at high temperature. Our retrieval solutions incorporate
ASSURE™ technology that contains a color-coded pH indicator for quality control and visual inspection.
Steam Monitor Strips
The heat and pressure sensitive steam strip allows the end-user to monitor both pressure and heat inside the Decloaking Chamber. The
best temperature for a Steam Monitor Strip to show quality control is ± 125 °C. It is at this temperature and relative pressure that a
dark brown to black color is achieved. Lower temperature and pressure will produce a lighter strip. The steam strip can be dated and
recorded for laboratory and CAP inspections. Cat. No. 613 (C,D,H)
QC Heat-Tested pH Strips
The pH of retrieval solutions at high temperatures can be critical for proper IHC staining. However, the pH of retrieval solutions can
change at high temperatures. To assure quality control and consistent staining, Biocare Medical has developed heat-tested strips that
give accurate and true pH at high temperatures. The pH of a solution at a certain temperature can be recorded on a daily basis to
comply with JCOHA and CAP recommendations. Cat. No. PH615 (ph7.5-10.5), PH616 (pH4-7)
Hot Sticker
Gasket
Removable Pan
Hot Sticker
Hot sticker will show this warning
when instrument is hot.
NOTE
: Be sure to remove any packing material from inside of the Decloaking Chamber before operating!
Failure to do so will result in FIRE.