G.7.4. « Maintenance » Parameter
→
Operating time: Indicates the total operating time of the UV lamp in hours (over the 13000h of lamp life). If the
ballast is faulty (the "UV lamp pb
” alarm is active), this time does not increase because the lamp is inactive or
disconnected.
Menu
Parameter
Function
Auxiliary
Maintenance
Total
Operating time: Indicates the total operating time of the UV lamp in hours
(over the 13000h of lamp life). If the ballast is faulty (the "UV lamp pb
” alarm
is active), this time does not increase because the lamp is inactive or
disconnected.
Reset
Resetting the total operating time counter to zero if the UV lamp is changed.
→
After 12500h, a "Change UV Lamp" information is displayed on the main screen
G.8. Information Menu
Menu
Meaning
Software Version
MASTER: XX.XX.XX
Control board program
Software Version
SLAVE: XX.XX.XX
Power card program
ID Code:
XXXXXXXX
Configuration code
Serial number:
XXXX-XXXXXX-XXX
Serial number
MAC Adress:
XXXXXXXXXXXX
MAC address for Bluetooth connection
MCU Temperature:
XX°C
Internal temperature in the electronics unit