Additional Specifications

Object temperature:

 0°C- 93.2°C (32°F-199.7°F)

Resolution ratio: 

0.1°C/°F

Measurement Location: 

laboratory

Accuracy: 

(35.0°C~42.0°C) ±0.2°C, (95.0°F~107.9°F) ±0.4°F, other 

temperature ±0.3°C

Relative maximum humidity:

 15%RH~93%RH

Atmospheric pressure: 

70kPa~106kPa

Relative maximum humidity: 

0%RH~93%RH

Atmospheric pressure: 

50kPa~106kPa

Display screen: 

LCD display screen, 4 bit numbers and icons.

1602 Forehead mode

Clinical bias, Dcb: 

0.078

Limits of Agreement, LA:

 .243

Clinical Repeatability, 

σ

r: 

0.069

The reference body site:

 forehead

Measuring site:

 forehead

Guidance and manufacturer’s declaration - electromagnetic emissions

RF emissions 
CISPR 11

The bt-degree IR Infrared Thermometer is intended for use in the 
electromagnetic environment specified below. The customer or the user 
of the bt-degree IR Infrared Thermometer should assure that it is used 

in such an environment.

Compliance

Electromagnetic
environment-guidance

Emissions

Group 1

The bt-degree IR Infrared 
Thermometer uses RF energy 
only for its internal function. 
Therefore, its RF emissions 
are very low and are not likely 
to cause any interference in 
nearby electronic equipment.

RF emissions 
CISPR 11

Class B

The bt-degree IR Infrared 
Thermometer is used in home 
and is powered by DC 3V

Harmonic emissions 
IEC 61000-3-2

N.A.

Voltage fluctuations/
flicker emissions IEC 
61000-3-3

N.A.

Guidance and manufacturer’s declaration – electromagnetic immunity

Electrostatic 

discharge 

(ESD)IEC 

61000-4-2

The bt-degree IR Infrared Thermometer is intended for use in the 

electromagnetic environment specified below. The customer or the user 

of the bt-degree IR Infrared Thermometer should assure that it is used 

in such an environment.

IEC 60601

test level

Electromagnetic

environment-guidance

Immunity

test

±8 kV 

contact 

±2 kV, ±4 

kV, ±8 kV, 

±15KV air

Floors should be wood, 

concrete or ceramic tile. 

If floors are covered with 

synthetic material, the 

relative humidity should be 

at least 30%.

Compliance

level 

±8 kV 

contact 

±2 kV, ±4 

kV, ±8 kV, 

±15KV air

Appendix 1 Guidance and Manufacturer Declaration Tables 

9

Содержание bt-degree IR

Страница 1: ...touchless infrared thermometer operation manual ...

Страница 2: ......

Страница 3: ...cifications Equipment Diagram Battery Installation Operation and Use Maintenance and Cleaning Troubleshooting Important Safety Precautions Disposal Environmental Conditions Definition of Symbols Warranty Appendix 1 2 2 2 3 3 5 6 7 7 7 8 8 9 ...

Страница 4: ... purchase of the bt degree IR Your equipment has been built with the highest quality components for lasting precision and operating durability IMPORTANT READ ALL INSTRUCTIONS BEFORE USE SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE 1 ...

Страница 5: ...speed Within 1 second Sounds Ready to measure measurement finished fever alert system error or fault audible on off function Fever alert 10 short fast beeps Memory 9 memory settings Automatic shutdown Unit will power off after 30 seconds of non use Water resistant IP22 Weight 2 4oz 68g without batteries Power supply 2 AAA batteries included Temperature units Fahrenheit F and Celsius C measurement ...

Страница 6: ...tery cover Operation and Use Sound Options 1 With the unit powered on press the mem button to select sound on or off A short beep will be audible and the symbol will appear 2 Press the mem button again the symbol will appear and the sound will be off Fahrenheit and Celsius setting options With power off press and hold the mem button for 6 seconds to switch between Fahrenheit settings and Celsius s...

Страница 7: ... 98 96 F 100 58 F 37 2 C 38 1 C 93 2 F 98 8 F 34 C 37 1 C 100 76 F 109 4 F 38 2 C 43 0 C Temperature readings 1 Press the button The LCD will display the measurement of the last reading A short beep will be heard when the unit is ready to measure 2 Hold and aim the sensor 1 18 2 3 5cm away from the center of the forehead 3 Press the button Hold the unit in place until you hear a long beep 4 The re...

Страница 8: ... 3 years No maintenance is required under normal conditions of use Contact the manufacturer or your sales representative with any questions regarding use or repair of equipment Cleaning To ensure precise measurements clean the bt degree IR after each use 1 Use a cotton swab with 75 alcohol to gently wipe the thermometer probe Wait 10 minutes before using the device after cleaning 2 To avoid damage...

Страница 9: ...9 4 F 43 0 C in the forehead temperature mode under 32 F 0 C in object temperature mode the LCD screen will show LO LO Er H The maximum operating environment temperature is 104 F If the environment temperature exceeds this point an error message will occur Operate in an environmental temperature of less than 104 F Er L The lower limit of operating environment temperature is 50 F If the environment...

Страница 10: ...o not open the unit There are no serviceable parts 8 Unauthorized modification or repair of the bt degree IR including the use of non original supplied parts or accessories can lead to injury damage failure and degradation of equipment performance 9 Do not operate if any portion of the bt degree IR is damaged in any way 10 Do not rinse or submerge under water or allow any liquids to enter the unit...

Страница 11: ...e and address Authorized Representative in the European Community Bio Therapeutic Micro Current Technology Inc guarantees the bt degree IR for 1 year after the date of purchase Defects due to faulty materials and workmanship will be repaired or replaced by Bio Therapeutic Micro Current Technology Inc provided that convincing proof of purchase in the qualifying period is provided Bio Therapeutic Mi...

Страница 12: ...idance Emissions Group 1 The bt degree IR Infrared Thermometer uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The bt degree IR Infrared Thermometer is used in home and is powered by DC 3V Harmonic emissions IEC 61000 3 2 N A Voltage fluctuations flicker e...

Страница 13: ... such an environment IEC 60601 test level Electromagnetic environment guidance 3 Vrms 150kHz to 80MHz 6 Vrms 150 kHz to 80 MHz outside ISM Portable and mobile RF communications equipment should be used no closer to any part of the bt degree IR Infrared Thermometer including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter C...

Страница 14: ... in the location in which the bt degree IR Infrared Thermometer is used exceeds the applicable RF compliance level above the bt degree IR Infrared Thermometer should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as re orienting or relocating the bt degree IR Infrared Thermometer d Over the frequency range 150 kHz to 80 MHz fiel...

Страница 15: ......

Страница 16: ... all countries The bt degree IR is not intended in the diagnosis of disease or other conditions or in the cure mitigation treatment or prevention of disease and is not intended to affect the structure of the human body biotherapeutic SA TM 1070 REV042320 Bio Therapeutic All rights reserved bio therapeutic com 206 938 5800 0197 ...

Отзывы: