TROUBLESHOOTING:

TECHNICAL FEATURES:

To contact Bio SB Technical Support

Phone: 1-800-561-1145

Email: [email protected]

INSTRUCTION MANUAL

TINTORETRIEVER PRESSURE COOKER 

For Epitope & Nucleic Acid Retrieval

Bio SB Inc.

 5385 Hollister Ave., Building 8, #108, Santa Barbara, CA 93111, USA

Tel (USA): 1-800-561-1145 | Tel (Int): +1-805-692-2768 | Fax: 805-692-2769

E-mail: [email protected]  |  Website: www.biosb.com

Cat. #: BSB-7087

Rated Capacity: 6L of Fluid or 96 Slides

Rated Power: 1000W

Rated Voltage: 110-220V 50-60Hz

Working Pressure: 0-60Kpa

Limited Pressure: 80Kpa

No.

Description

Cause

Solution

1

Unable to close lid.

The installation of the aluminum lid 
with gasket is not correct.

Properly position the aluminum lid with gasket 
on support.

The float valve is jammed.

Gently poke the float valve with a paper clip or 
tooth pick.

2

Unable to open the 
pressure cooker lid.

The float valve doesn’t fall.

Use a paper clip or toothpick to lower the float 
valve.

3

Excess vapor leaking from 
pressure cooker lid.

The installation of the sealing gasket 
is not correct.

Properly set the sealing gasket on the aluminum 
lid.

Sealing gasket has residue.

Clean the sealing gasket.

Sealing gasket is damaged.

Replace the sealing gasket.

The pressure cooker lid is not properly 
closed.

Cover the lid according to the manual.

4

Air leaks from the float 
valve.

The gasket of the float valve has 
residue or particles.

Clean the gasket of the float valve.

The gasket of the float valve is 
damaged.

Replace the gasket of the float valve.

5

The float valve is not rising. Internal pressure doesn’t reach the 

minimum limit.

Contact Technical Support for assistance.

Air leakage of the pressure cooker lid or 
the discharge pressure valve.

Contact Technical Support for assistance.

6

E1

Open circuit of the sensor.

Contact Technical Support for assistance.

7

E2

Short circuit of the sensor.

Contact Technical Support for assistance.

8

E3

Overheating.

Contact Technical Support for assistance.

9

E4

Pressure Switch Failure.

Contact Technical Support for assistance.