1.3 Safety
The Model GS-800 Calibrated Imaging Densitometer system uses high voltage and high
current and should be operated with care at all times. To avoid shock, set up the GS-800
Densitometer in a dry area. Immediately wipe up any spilled solutions.
• Turn off the Densitometer, and then disconnect the power plug when you want to
clean the Densitometer case or glass plate, or when the Densitometer needs service or
repair.
• Place the Densitometer on a level surface.
• To ensure proper ventilation, allow a minimum of 15 cm free space around each side of
the Densitometer.
• Do not leave photographs, gels, or film on the glass plate for excessive periods of time.
The heat from the light source in the Densitometer may cause them to deteriorate.
• Do not operate the Densitometer when the environmental temperature falls below 5° C
or rises above 40° C.
• Do not operate the Densitometer when the environmental humidity falls below 25% or
rises above 85%.