BIO PROTECH MAXEMS 1000 Скачать руководство пользователя страница 9

14

g. Set Timer
    The treatment time can be adjusted 15 minutes, 30 minutes, 45 minutes,
    60 minutes.

9. Check/Replace the Battery: 
    Over time, in order to ensure the functional safety of
    EMS, changing the battery is necessary.
    1) Make sure that both intensity controls are switched
        to off-position.
    2) Slide the battery compartment cover and remove.
    3) Remove the battery from the compartment.
    4) Insert the battery into the compartment. Note the 
        polarity indicated on the battery and in the compartment.
    5) Replace the battery compartment cover and slide
        to close.
    6) If the battery symbol on the LCD shows “Low” battery, 
        replace battery with new one.

                         Chapter 18: INSTRUCTION FOR USE

Protocol for the safe application of EMS

1) Check contraindications with patient.

2) Test skin for normal sensation using blunt/sharp test.

3) EMS device should be switched off and electrode leads disconnected.

4) Set electrical characteristics of EMS while device is switched off. They

   should be prescribed.

5) Connect electrodes to pins on lead wire and position electrodes on patient’s skin.

6) Ensure EMS device is still switched off and connect the electrode wire to the

    EMS device.

7) Switch the EMS device ON.

15

Select Menu

Menu Up

Menu Down

8) Gradually (slowly) increase the intensity until the patient experiences the first 
    ‘tingling’ sensation from the stimulator.
9) Gradually (slowly) increase the intensity further until the patient experiences 
    a ‘strong but comfortable’ tingling sensation.
10) This intensity should not be painful.

Protocol for the safe termination of EMS

1) Gradually (slowly) decrease the intensity until the patient experiences no 
    tingling sensation.
2) Switch the EMS device OFF.
3) Disconnect the electrode wire from the EMS device.
4) Disconnect electrodes from the pins on lead wire.
5) Remove the electrodes from the patient’s skin.

              Chapter 19: MAINTENANCE, TRANSPORTATION 

                                AND STORAGE OF MAXEMS 1000

1. Non-flammable cleaning solution is suitable for cleaning the device.
    Note: Do not smoke or work with open flames (for example, candles, etc.) 
             when working with flammable liquids
2. Stains and spots can be removed with a cleaning agent.
3. Do not submerge the device in liquids or expose it to large amounts of water.
4. Return the device to the carrying box with sponge foam to ensure that the unit
    is well-protected before transportation.
5. If the device is not to be used for a long period of time, remove the batteries
    from the battery compartment (acid may leak from used batteries and damage 
    the device). Put the device and accessories in carrying box and keep it in a cool
    dry place.
6. The packed MAXEMS 1000 should be stored and transported under the

    temperature range of 

-10℃~40℃(14℉ ~104℉)

, relative humidity 30% ~ 75%,

    atmosphere pressure 700 hPa ~ 1060hPa.

                  Chapter 20: SAFETY-TECHNICAL CONTROLS

For safety reasons, check your MAXEMS 1000 each week based on the following 
checklist.
1. Check the device for external damage.
    - Deformation of the housing.
    - Damaged or defective output sockets.

Содержание MAXEMS 1000

Страница 1: ...NSTRUCTION MANUAL for the MAXEMS 2000 Manufactured by Bio Protech Inc Donghwa Medical Instrument complex 151 3 Donghwagongdan ro Munmak eup Wonju si Gangwon do 26365 Korea Tel 82 33 735 7720 Fax 82 33...

Страница 2: ...ECIFICATION 9 ACCESSORIES 10 GRAPHIC SYMBOLS 11 ATTACHMENT OF ELECTRODE LEAD WIRES 12 LEAD WIRE MAINTENANCE 13 ELECTRODE OPTIONS 14 ELECTRODE PLACEMENT 15 TIP FOR SKIN CARE 16 APPLICATION OF RE USABLE...

Страница 3: ...erating machinery or during any activity in which involuntary muscle contractions may put the user at undue risk of injury Chapter 5 ADVERSE REACTIONS Skin irritation and burns beneath the electrodes...

Страница 4: ...2 Pulse Amplitude Adjustable by intensity control knob Adjustable 0 80mA peak into 500 ohm load each channel 3 Output Voltage Adjustable by intensity control knob Adjustable 0 40V peak into 500 ohm l...

Страница 5: ...r your condition Follow application procedures outlined in electrode packaging to maintain stimulation and prevent skin irritation Use of legally marketed adhesive EMS electrodes is recommended EMS el...

Страница 6: ...lling stress tape extra lengths of lead wires to the skin in a loop to prevent tugging on electrodes 6 When removing electrodes always remove by pulling in the direction of hair growth 7 It may be hel...

Страница 7: ...must be switched off before connecting the cables Both intensity controls must be at the off position Electrodes must be pressed firmly on the skin 4 Function Selector Expose the controls by sliding f...

Страница 8: ...setting value to 0 When the ramp time is set each contraction may be ramped in order that signals come on and come off gradually and smoothly The ramp time is adjustable from 1 second to 8 seconds by...

Страница 9: ...her until the patient experiences a strong but comfortable tingling sensation 10 This intensity should not be painful Protocol for the safe termination of EMS 1 Gradually slowly decrease the intensity...

Страница 10: ...essary insert a new battery 4 For possible damage to the cable Change the cable if any damage is detected If there is any other problem please return the device to your distributor Do not try to repai...

Отзывы: