20
IV. SET UP
CAUTION: Antistatic or electrically conductive hoses or tubing should not be used.
A. CONNECTION OF GAS SUPPLY
1. Connect a 50 psi (345 kPa) oxygen source to the logic power gas supply
connector. Only 100% oxygen should be used for proper operation.
2. Connect the patient gas supply to the DISS fitting labeled "patient gas
supply." This may be from the output of a blender or any pre-blended gas
mixture of the desired oxygen concentration. The oxygen concentration
delivered to the patient is the same as the oxygen concentration supplied to
the ventilator in the patient gas fitting.
It is essential that:
THE SUPPLIES ARE REGULATED 50 PSI (345 kPa) SOURCES
WITHOUT FLOW RESTRICTING DEVICES (E.G., FLOWMETER,
NEEDLE VALVE, ETC.);
HOSE FITTINGS SHOULD BE HAND TIGHTENED TO AVOID DAMAGE
TO FITTINGS;
THE GAS SUPPLY SHOULD BE CLEAN AND DRY.
B. CONNECTION OF PATIENT CIRCUIT
1. Connect the corrugated humidifier hose to the main patient output fitting.
Connect the other end to the input port of the humidifier.
2. Connect the main patient hose (corrugated tube) to the output connector of
the humidifier.
3. Attach the exhalation valve line to the exhalation valve fitting.
4. Connect the proximal airway pressure line to the pressure gauge fitting.
5. Attach patient port of exhalation valve to a test lung, BMD Part #1020. After
selecting desired operating parameters observe proper functioning before
attaching to the patient.
CAUTION:
Do not re-use disposable breathing circuits.
Note:
Any intensive care ventilator circuit can be used with the IC-2A including a
configuration utilizing both inspiratory and expiratory hoses. Any exhalation valve
may be used. It should be noted that the IC-2A is supplied with the Bio-Med
Devices patient circuit, catalog #8002A. The maximum pressure limit and maximum
PEEP pressure are calibrated using this exhalation valve. If another valve is used
there may be a difference in the maximum pressure limit and maximum PEEP
pressure attainable, depending on the area ratios of the exhalation valve used.