7. Maintenance and Storage
Exterior Pump Case Cleaning instructions:
1. Clean the exterior case and tubing with a damp (not wet) cloth using mild soap and water
solution once per month.
WARNING!
• Only an authorized technician may open the pump
• Before cleaning, unplug power cord from electrical outlet
2. Disconnect garment from device.
3. Open garment to expose all sides either by separating Velcro type hook and loop or by
unzipping (depending on type of garment).
WARNING!
•
Do not allow liquids to enter pump, as this can present an electrical hazard
•
Always allow the pump to dry before using
•
Do not use bleach on the pump
4. Wipe the garments thoroughly with a ph neutral cleaner or Clinell wipes. Solutions with
alcohol should be avoided.
Garment cleaning/disinfecting instructions:
WARNING!
• Never allow the Latch Connectors to be submerged into the water. If water enters the
inside of the garment, damage may occur to the device.
• DO NOT place garment in washing machine.
• WARNING! DO NOT use the tubing or valves as “handles” for carrying, handing or
storing garment.
The unit contains no user serviceable parts and should only be serviced by a competent
electrical technician or returned to Medi-Rent Pty Ltd or your local authorsied dealer.
All Biocompression products should be serviced regularly by Medi-Rent Pty Ltd or an
authorsied dealer in order to comply with warranty conditions.
NB. Before returning equipment to Medi-Rent Pty Ltd for service, be sure that it has been
properly cleaned and disinfected in accordance with local health service guidelines.
8. Specifications
Dimensions
11.43 H x 29.86 W
x 19.69 D cm
Weight
2.5kg
Inflation/Deflation
44/5.5 Seconds
Cycle Time
5.5 Secs / chamber
Electrical
120 VAC, 60 Hz, 0.5
A (USA) 230 VAC,
Applied Part
Type B
Protection against
Electrical Shock
CLASS II (USA)
CLASS IIa (EU)
Operation Mode
Continuous
Operation
General Equipment Specifications
Environmental Conditions
For Operation
Ambient temperature:
+10°C - +44°C
Relative humidity:
30% - 75%
Atmospheric pressure:
700hPa to 1060hPa
For Transport and Storage
Ambient temperature:
+10°C - +44°C
Relative humidity:
30% - 75%
Atmospheric pressure:
700hPa to 1060hPa
Electrical Specifications/Equipment Classification
The Sequential Circulator SC-2008-OC interior components are “double insulated” and do not
require a “protective ground.” The devices are equipped with 18 gauge, 2-wire, 10’ Power Cords,
secured through the pump casings with a Heyco strain relief brushing, as well as, an additional
“hold-down” clamp for added safety.
• Class of protection against electrical shock: CLASS II EQUIPMENT
• The degree of protection against electric shock: APPLIED PART-TYPE B
• Mode: CONTINUOUS OPERATION WITH INTERMITTENT LOADING
• According degree of protection against ingress of water: IXP0
16