7. Maintenance and Storage

Exterior Pump Case Cleaning instructions:

1.  Clean the exterior case and tubing with a damp (not wet) cloth using mild soap and water 

solution once per month.

WARNING! 

•  Only an authorized technician may open the pump 
•  Before cleaning, unplug power cord from electrical outlet

 

 

2.  Disconnect garment from device. 

3.  Open garment to expose all sides either by separating Velcro type hook and loop or by 

unzipping (depending on type of garment).

WARNING! 

• 

Do not allow liquids to enter pump, as this can present an electrical hazard

• 

Always allow the pump to dry before using 

• 

Do not use bleach on the pump 

4.  Wipe the garments thoroughly with a ph neutral cleaner or Clinell wipes. Solutions with 

alcohol should be avoided.

Garment cleaning/disinfecting instructions:

WARNING! 

•  Never allow the Latch Connectors to be submerged into the water. If water enters the 

inside of the garment, damage may occur to the device.

•  DO NOT place garment in washing machine.
•  WARNING! DO NOT use the tubing or valves as “handles” for carrying, handing or 

storing garment.

 
The unit contains no user serviceable parts and should only be serviced by a competent 
electrical technician or returned to Medi-Rent Pty Ltd or your local authorsied dealer.

All Biocompression products should be serviced regularly by Medi-Rent Pty Ltd or an 
authorsied dealer in order to comply with warranty conditions.

NB. Before returning equipment to Medi-Rent Pty Ltd for service, be sure that it has been 
properly cleaned and disinfected in accordance with local health service guidelines. 

 

8. Specifications 

Dimensions

11.43 H  x 29.86  W  
x  19.69 D cm 

Weight

2.5kg

Inflation/Deflation

44/5.5 Seconds

Cycle Time

5.5 Secs / chamber

Electrical

120 VAC, 60 Hz, 0.5 
A (USA) 230 VAC,

Applied Part

Type B

Protection against 
Electrical Shock

CLASS II (USA) 
CLASS IIa (EU)

Operation Mode

Continuous 
Operation

General Equipment Specifications

Environmental Conditions

For Operation

Ambient temperature:

+10°C -  +44°C

Relative humidity:

30% - 75%

Atmospheric pressure:

700hPa to 1060hPa

For Transport and Storage

Ambient temperature:

+10°C -  +44°C

Relative humidity:

30% - 75%

Atmospheric pressure:

700hPa to 1060hPa

Electrical Specifications/Equipment Classification

The Sequential Circulator SC-2008-OC interior components are “double insulated” and do not 
require a “protective ground.” The devices are equipped with 18 gauge, 2-wire, 10’ Power Cords, 
secured through the pump casings with a Heyco strain relief brushing, as well as, an additional 
“hold-down” clamp for added safety.  

•  Class of protection against electrical shock: CLASS II EQUIPMENT 
•  The degree of protection against electric shock: APPLIED PART-TYPE B 
•  Mode: CONTINUOUS OPERATION WITH INTERMITTENT LOADING 
•  According degree of protection against ingress of water: IXP0

16

17

Содержание SC 2008 OC

Страница 1: ...Indications for safety use Biocompression PumpSC2008OC UserManual Indications for safe use This product is intended for use by health professionals or patients who are under clinical supervision...

Страница 2: ...ulators SC 2008 OC are pneumatic compression devices intended for either primary or adjunctive treatment of primary or secondary lymphoedema and lipoedema The devices are also intended for additional...

Страница 3: ...do not stand or walk while wearing garments Precautions This device is not intended for use during SLEEP Do not attempt to modify this device in any way Contraindications Compression is NOT recommend...

Страница 4: ...Blocker Bars for use during single garment therapy 1 Power Cord Necessary garments ordered LATCH CONNECTORS Attach and detach the garment to the pump via the Receptor Ports and Air supply ports depic...

Страница 5: ...Vest 3 Product description Garments Accessories 16 Chamber Pants 8 Chamber Legs 3 Product description Garments Optional Accessories Your product may have the following accessories Carry bag additiona...

Страница 6: ...thout rotating twisting or large creasing of the material Medi Rent Pty Limited ABN 37 603 454 360 Unit 3 61 Beauchamp Rd Matraville NSW 2036 P 1300 726 666 F 02 9316 9998 E info medirent com au W www...

Страница 7: ...ther device other than the Biocompression Controller Never expose garments to oil benzene alcohol hydrocarbons etc For cleaning refer to the Maintenance and Storage section of this manual Remove jewel...

Страница 8: ...presenting condition Stop the device during operation Recommendation 6 Troubleshooting Symptom Possible Cause Corrective Action The device is not working No electricity Check the electrical wall outl...

Страница 9: ...d be serviced regularly by Medi Rent Pty Ltd or an authorsied dealer in order to comply with warranty conditions NB Before returning equipment to Medi Rent Pty Ltd for service be sure that it has been...

Страница 10: ...IEC 61000 3 2 Not applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Recommended separation distances between portable and mobile RF communications equipment and the SC 20...

Страница 11: ...h minimal inconvenience For this reason it is important that you obtain a Return Material Authorization Number RMA when calling Customer Service You can feel confident that your product is backed by t...

Страница 12: ...MEDIRENT COM AU 2020 22 23...

Страница 13: ...t 3 61 Beauchamp Rd Matraville 2067 NSW Ph 1300 726 666 info medirent com au www medirent com au 2020 MEDI RENT Unit 3 61 Beauchamp Rd Matraville 2067 NSW Ph AU 02 9700 1744 Ph NZ 09 972 3068 info med...

Отзывы: