Information for Distributors and Healthcare Providers
Compliance Meter
To see hours of use since last reset, refer to “To See Treatment
Compliance Meter (Hours of Use)” in the Operating Instructions section.
To see pump total hours of use, press “Play/Pause” for 10 seconds while
viewing hours since last reset -
“t” will appear followed by the total hours of use.
Resetting the Pump
The pump remembers user settings and therefore it is important to reset the
pump to its original factory settings when placing the device on a new patient.
To reset the treatment compliance meter and return the pump to factory
settings:
Begin with the device turned off and plugged in
Press and hold the UP, DOWN, and SET buttons simultaneously for 5
seconds
LED display will light up and the buttons can be released
Press “SET” button – LED display will turn off and pump has been reset